MP13: Association between the quantity of subcutaneous fat and the inter-device agreement of two tissue oximeters. (May 2019)
- Record Type:
- Journal Article
- Title:
- MP13: Association between the quantity of subcutaneous fat and the inter-device agreement of two tissue oximeters. (May 2019)
- Main Title:
- MP13: Association between the quantity of subcutaneous fat and the inter-device agreement of two tissue oximeters
- Authors:
- Cournoyer, A.
Cossette, S.
Paquet, J.
Daoust, R.
Marquis, M.
Notebaert, É.
Iseppon, M.
Chauny, J.
Denault, A. - Abstract:
- Abstract : Introduction : Near-infrared spectroscopy (NIRS) can be used to monitor the oxygen saturation of hemoglobin in any given superficial tissue. However, the measurements provided by different oximeters can vary a lot. Little is known about the specific patient characteristics that could affect the inter-device agreement of tissular oximeters. This study aimed to evaluate the association between the quantity of subcutaneous fat (assessed by skinfold thickness) and the inter-device agreement of two tissue oximeters, the INVOS 5100c and the Equanox 7600.Methods : In this prospective cohort study, tissue saturations and skinfold thickness were measured at four different sites on both sides of the body in healthy adult (≥18 years old) volunteers. The association between the quantity of subcutaneous fat (assessed by skinfold thickness) and the inter-device agreement (absolute difference between the oximetry values provided by the two oximeters) was first assessed with a Pearson's correlation and a scatter plot. Subsequently, a linear mixed model was used to evaluate the impact of the subcutaneous fat and other covariables (age, sex) on the inter-device agreement while adjusting for the repeated measurements across different sites for the same volunteers.Results : From January to March 2015, 53 healthy volunteers were included in this study with ages ranging between 20 and 81 years old, on which a total of 848 measures were taken. Higher skinfold measures were associatedAbstract : Introduction : Near-infrared spectroscopy (NIRS) can be used to monitor the oxygen saturation of hemoglobin in any given superficial tissue. However, the measurements provided by different oximeters can vary a lot. Little is known about the specific patient characteristics that could affect the inter-device agreement of tissular oximeters. This study aimed to evaluate the association between the quantity of subcutaneous fat (assessed by skinfold thickness) and the inter-device agreement of two tissue oximeters, the INVOS 5100c and the Equanox 7600.Methods : In this prospective cohort study, tissue saturations and skinfold thickness were measured at four different sites on both sides of the body in healthy adult (≥18 years old) volunteers. The association between the quantity of subcutaneous fat (assessed by skinfold thickness) and the inter-device agreement (absolute difference between the oximetry values provided by the two oximeters) was first assessed with a Pearson's correlation and a scatter plot. Subsequently, a linear mixed model was used to evaluate the impact of the subcutaneous fat and other covariables (age, sex) on the inter-device agreement while adjusting for the repeated measurements across different sites for the same volunteers.Results : From January to March 2015, 53 healthy volunteers were included in this study with ages ranging between 20 and 81 years old, on which a total of 848 measures were taken. Higher skinfold measures were associated with an increase in the difference between measures provided by both oximeters (Slope = -0.59, Pearson correlation coefficient = -0.51, p < 0.001). This observed association persisted in a linear mixed model (-0.48 [95% confidence interval {CI}-0.61 to -0.36], p < 0.001). The sex of the volunteers also influenced the inter-oximeter agreement (Women:-5.77 [95%CI -8.43 to -3.11], p < 0.001), as well as the forearm sites (Left forearm: −7.16 [95%CI -9.85 to –4.47], p < 0.001; right forearm: −7.01 [95%CI -9.61 to −4.40], p < 0.001).Conclusion : The quantity of subcutaneous fat, as well as the sex of the volunteers and the measurement sites, impacted the inter-device agreement of two commonly used oximeters. Given these findings, monitoring using tissue oximetry should be interpreted with great care when there is a significant quantity of subcutaneous fat. … (more)
- Is Part Of:
- CJEM. Volume 21(2019)Supplement 1
- Journal:
- CJEM
- Issue:
- Volume 21(2019)Supplement 1
- Issue Display:
- Volume 21, Issue 1 (2019)
- Year:
- 2019
- Volume:
- 21
- Issue:
- 1
- Issue Sort Value:
- 2019-0021-0001-0000
- Page Start:
- S46
- Page End:
- S47
- Publication Date:
- 2019-05
- Subjects:
- inter-device agreement, -- near-infrared spectroscopy, -- tissular oximetry
Emergency Treatment -- Periodicals
Emergency Medicine -- Periodicals
Emergency medical services -- Canada -- Periodicals
Medical emergencies -- Canada -- Periodicals
Emergency medical services
Medical emergencies
Canada
Periodicals
616.02505 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=CEM ↗
http://www.caep.ca/004.cjem-jcmu/004-00.cjem/004-01v.archives.htm#main ↗
http://link.springer.com/ ↗ - DOI:
- 10.1017/cem.2019.148 ↗
- Languages:
- English
- ISSNs:
- 1481-8035
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 11057.xml