Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate. Issue 7 (5th April 2023)
- Record Type:
- Journal Article
- Title:
- Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate. Issue 7 (5th April 2023)
- Main Title:
- Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO2, VCO2, and resting metabolic rate
- Authors:
- Henriksen, Hege Berg
Henriksen, Christine
de Sousa, Ana Rita Sequeira
Alavi, Dena Treider
Augestad, Elin Maria Sandstad
Rising, Russell
Dörner, Rebecca
Hägele, Franziska Anna
Bosy‐Westphal, Anja
Blomhoff, Rune
Ulven, Stine Marie
Olsen, Thomas - Abstract:
- Abstract: Whole‐room indirect calorimeters (WRICs) provide accurate instruments for the measurement of respiratory exchange, energy expenditure, and macronutrient oxidation. Here, we aimed to determine the validity and reproducibility of a 7500 L WRIC for the measurement of ventilation rates and resting metabolic rate (RMR). Technical validation was performed with propane combustion tests ( n = 10) whereas biological reproducibility was tested in healthy subjects (13 women, 6 men, mean ± SD age 39.6 ± 15.3) in two 60 min measurements separated by 24 h. Subjects followed a run‐in protocol prior to measurements. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were calculated for ventilation rates of O2 (VO2), CO2 (VCO2), the respiratory quotient (RQ; VCO2/VO2), and RMR. Technical validation showed good validity with CVs ranging from 0.67% for VO2 to 1.00% for energy expenditure. For biological reproducibility, CVs were 2.89% for VO2 ; 2.67% for VCO2 ; 1.95% for RQ; and 2.68% for RMR. With the exception of RQ (74%), ICCs were excellent for VO2 (94%), VCO2 (96%) and RMR (95%). Excluding participants that deviated from the run‐in protocol did not alter results. In conclusion, the 7500 L WRIC is technically valid and reproducible for ventilation rates and RMR. Abstract : Whole‐room indirect calorimetry is an accurate method for measuring metabolic rates. Here, we report that a newly built whole‐room indirect calorimeter at the Department ofAbstract: Whole‐room indirect calorimeters (WRICs) provide accurate instruments for the measurement of respiratory exchange, energy expenditure, and macronutrient oxidation. Here, we aimed to determine the validity and reproducibility of a 7500 L WRIC for the measurement of ventilation rates and resting metabolic rate (RMR). Technical validation was performed with propane combustion tests ( n = 10) whereas biological reproducibility was tested in healthy subjects (13 women, 6 men, mean ± SD age 39.6 ± 15.3) in two 60 min measurements separated by 24 h. Subjects followed a run‐in protocol prior to measurements. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were calculated for ventilation rates of O2 (VO2), CO2 (VCO2), the respiratory quotient (RQ; VCO2/VO2), and RMR. Technical validation showed good validity with CVs ranging from 0.67% for VO2 to 1.00% for energy expenditure. For biological reproducibility, CVs were 2.89% for VO2 ; 2.67% for VCO2 ; 1.95% for RQ; and 2.68% for RMR. With the exception of RQ (74%), ICCs were excellent for VO2 (94%), VCO2 (96%) and RMR (95%). Excluding participants that deviated from the run‐in protocol did not alter results. In conclusion, the 7500 L WRIC is technically valid and reproducible for ventilation rates and RMR. Abstract : Whole‐room indirect calorimetry is an accurate method for measuring metabolic rates. Here, we report that a newly built whole‐room indirect calorimeter at the Department of Nutrition, University of Oslo provides valid and reproducible measurements of resting metabolic rate. … (more)
- Is Part Of:
- Physiological reports. Volume 11:Issue 7(2023)
- Journal:
- Physiological reports
- Issue:
- Volume 11:Issue 7(2023)
- Issue Display:
- Volume 11, Issue 7 (2023)
- Year:
- 2023
- Volume:
- 11
- Issue:
- 7
- Issue Sort Value:
- 2023-0011-0007-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2023-04-05
- Subjects:
- energy expenditure -- metabolism -- resting metabolic rate -- whole‐room indirect calorimetry
Physiology -- Periodicals
571 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)2051-817X ↗
http://physreports.physiology.org ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.14814/phy2.15658 ↗
- Languages:
- English
- ISSNs:
- 2051-817X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 27086.xml