Validation of the ABPMpro ambulatory blood pressure monitor in the general population according to AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018). Issue 3 (5th June 2023)
- Record Type:
- Journal Article
- Title:
- Validation of the ABPMpro ambulatory blood pressure monitor in the general population according to AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018). Issue 3 (5th June 2023)
- Main Title:
- Validation of the ABPMpro ambulatory blood pressure monitor in the general population according to AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)
- Authors:
- Roth, Bernhard
Bothe, Tomas Lucca
Patzak, Andreas
Pilz, Niklas - Abstract:
- Abstract : Objective: The objective of this study is to evaluate the accuracy of the oscillometric upper-arm device ABPMpro (SOMNOmedics) for ambulatory blood pressure measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) at rest and during dynamic exercise. Methods: Subjects were recruited to fulfill the age, sex, blood pressure (BP) and cuff distribution criteria of the AAMI/ESH/ISO standard using the same arm sequential BP measurement method. Three appropriate cuff sizes (18–24, 24–34 and 34–46 cm) of the tested device were used for the arm-varying circumferences. The inflation and deflation measurement modes of the ABPMpro were investigated. Results: For the general validation study, 100 subjects were recruited and 90 were analyzed. For validation criterion (1), the mean ± SD of the differences between ABPMpro and reference BP was 0.7 ± 7.3/–0.7 ± 5.8 mmHg (systolic/diastolic) for inflation and 1.4 ± 7.7/–0.6 ± 6.1 mmHg for deflation measurements. For criterion (2), the SD of the averaged BP differences per subject was 5.98/5.10 mmHg for inflation and 6.46/5.36 mmHg for deflation measurements, thereby passing the threshold. In the ambulatory validation study ( N = 36), the mean difference was -1.2 ± 7.9/ 2.4 ± 6.6 mmHg for inflation and –0.7 ± 7.6/3.1 ± 7.0 mmHg for deflationAbstract : Objective: The objective of this study is to evaluate the accuracy of the oscillometric upper-arm device ABPMpro (SOMNOmedics) for ambulatory blood pressure measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) at rest and during dynamic exercise. Methods: Subjects were recruited to fulfill the age, sex, blood pressure (BP) and cuff distribution criteria of the AAMI/ESH/ISO standard using the same arm sequential BP measurement method. Three appropriate cuff sizes (18–24, 24–34 and 34–46 cm) of the tested device were used for the arm-varying circumferences. The inflation and deflation measurement modes of the ABPMpro were investigated. Results: For the general validation study, 100 subjects were recruited and 90 were analyzed. For validation criterion (1), the mean ± SD of the differences between ABPMpro and reference BP was 0.7 ± 7.3/–0.7 ± 5.8 mmHg (systolic/diastolic) for inflation and 1.4 ± 7.7/–0.6 ± 6.1 mmHg for deflation measurements. For criterion (2), the SD of the averaged BP differences per subject was 5.98/5.10 mmHg for inflation and 6.46/5.36 mmHg for deflation measurements, thereby passing the threshold. In the ambulatory validation study ( N = 36), the mean difference was -1.2 ± 7.9/ 2.4 ± 6.6 mmHg for inflation and –0.7 ± 7.6/3.1 ± 7.0 mmHg for deflation measurements. Conclusion: The ABPMpro device fulfilled the ISO 81060-2:2018 requirements in the general population and in the ambulatory setting and can therefore be recommended for clinical use. … (more)
- Is Part Of:
- Blood pressure monitoring. Volume 28:Issue 3(2023)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 28:Issue 3(2023)
- Issue Display:
- Volume 28, Issue 3 (2023)
- Year:
- 2023
- Volume:
- 28
- Issue:
- 3
- Issue Sort Value:
- 2023-0028-0003-0000
- Page Start:
- 158
- Page End:
- 162
- Publication Date:
- 2023-06-05
- Subjects:
- accuracy -- ambulatory blood pressure monitoring -- exercise -- ISO protocol -- validation
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000640 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 27060.xml