Biomechanical comparison between double-plate fixation and posterior plate fixation for comminuted olecranon fracture using two triceps screws in synthetic bone model. Issue 1 (7th January 2022)
- Record Type:
- Journal Article
- Title:
- Biomechanical comparison between double-plate fixation and posterior plate fixation for comminuted olecranon fracture using two triceps screws in synthetic bone model. Issue 1 (7th January 2022)
- Main Title:
- Biomechanical comparison between double-plate fixation and posterior plate fixation for comminuted olecranon fracture using two triceps screws in synthetic bone model
- Authors:
- Lee, Yohan
Cho, Bong Wan
Kim, Min Bom
Lee, Young Ho - Other Names:
- Dwivedi. Atul section editor.
- Abstract:
- Abstract: Background: Although preventing triceps fragment displacement is essential for treating an olecranon fracture, we frequently encounter situations in which only a few screws can be fixed to the triceps fragment. The aim of this study was to compare the stability of double-plate fixation and posterior plate fixation for olecranon fractures when the triceps fragment was small and only 2 screws could be inserted. Methods: A composite ulna model was used to simulate olecranon fracture. Four groups were formed consisting of double-plate and posterior plates with cortical and locking screws. The cyclic loading test was conducted for 500 cyclic loads of 5 to 50 N on a specimen to measure micromotion and displacement of the gap caused by light exercise. The load-to-failure test was performed by applying a load until fixation loss, defined as when the fracture gap increased by 2 mm or more or catastrophic failure occurred, to measure the maximum load. Results: Eight samples per group were tested through the pilot study. All groups were stable with a micromotion of <0.5 mm. However, the mean micromotion showed significant differences between the 4 groups ( P < .001, Table 1 ). In the mean micromotion during exercise, posterior plating with cortical screws was the most stable (0.09 ± 0.02 mm) while double-plating with cortical screws was the most unstable (0.42 ± 0.11 mm). At the maximum load, posterior plating with locking screws was the strongest (205.3 ± 2.8 N) whileAbstract: Background: Although preventing triceps fragment displacement is essential for treating an olecranon fracture, we frequently encounter situations in which only a few screws can be fixed to the triceps fragment. The aim of this study was to compare the stability of double-plate fixation and posterior plate fixation for olecranon fractures when the triceps fragment was small and only 2 screws could be inserted. Methods: A composite ulna model was used to simulate olecranon fracture. Four groups were formed consisting of double-plate and posterior plates with cortical and locking screws. The cyclic loading test was conducted for 500 cyclic loads of 5 to 50 N on a specimen to measure micromotion and displacement of the gap caused by light exercise. The load-to-failure test was performed by applying a load until fixation loss, defined as when the fracture gap increased by 2 mm or more or catastrophic failure occurred, to measure the maximum load. Results: Eight samples per group were tested through the pilot study. All groups were stable with a micromotion of <0.5 mm. However, the mean micromotion showed significant differences between the 4 groups ( P < .001, Table 1 ). In the mean micromotion during exercise, posterior plating with cortical screws was the most stable (0.09 ± 0.02 mm) while double-plating with cortical screws was the most unstable (0.42 ± 0.11 mm). At the maximum load, posterior plating with locking screws was the strongest (205.3 ± 2.8 N) while double-plating with cortical screws was the weakest (143.3 ± 27.1 N). There was no significant difference in displacement after light exercise between the groups. Conclusions: This study showed that when 2 triceps screws were used, both groups were stable during light exercise, but posterior-plating was stronger than double-plating. Abstract : Supplemental Digital Content is available in the text … (more)
- Is Part Of:
- Medicine. Volume 101:Issue 1(2021)
- Journal:
- Medicine
- Issue:
- Volume 101:Issue 1(2021)
- Issue Display:
- Volume 101, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 101
- Issue:
- 1
- Issue Sort Value:
- 2021-0101-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-01-07
- Subjects:
- biomechanical phenomena -- bone plates -- comminuted/surgery -- fractures -- olecranon process/surgery
Medicine -- Periodicals
Medicine -- Periodicals
Médecine -- Périodiques
Geneeskunde
Medicine
Periodicals
Periodicals
610.5 - Journal URLs:
- http://journals.lww.com/md-journal/pages/default.aspx ↗
http://gateway.ovid.com/ovidweb.cgi?T=JS&PAGE=toc&D=ovft&MODE=ovid&NEWS=N&AN=00002060-000000000-00000 ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MD.0000000000028313 ↗
- Languages:
- English
- ISSNs:
- 0025-7974
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5534.000000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 26999.xml