Detailed investigation of the lesion formation with a novel contact force sensing catheter with a mesh‐shaped irrigation tip. Issue 2 (27th February 2023)
- Record Type:
- Journal Article
- Title:
- Detailed investigation of the lesion formation with a novel contact force sensing catheter with a mesh‐shaped irrigation tip. Issue 2 (27th February 2023)
- Main Title:
- Detailed investigation of the lesion formation with a novel contact force sensing catheter with a mesh‐shaped irrigation tip
- Authors:
- Matsumoto, Kazuhisa
Kawano, Daisuke
Sasaki, Wataru
Tanaka, Naomichi
Narita, Masataka
Mori, Hitoshi
Tsutsui, Kenta
Ikeda, Yoshifumi
Arai, Takahide
Nakano, Shintaro
Kato, Ritsushi
Matsumoto, Kazuo - Abstract:
- Abstract: Background: Recently, a novel contact force (CF) sensing catheter with mesh‐shaped irrigation tip (TactiFlex SE, Abbott) was invented and is expected to be useful for safe and effective radiofrequency ablation. However, this catheter's detailed characteristics of the lesion formation are unknown. Methods: With an in vitro model, TactiFlex SE and its predecessor, FlexAbility SE, were used. A cross‐sectional analysis of 60 s lesions (combination of various energy power settings [30, 40, and 50 W], and CFs [10, 30, and 50 g]) and longitudinal analysis (combination of various powers [40 or 50 W], CFs [10, 30, and 50 g] and ablation times [10, 20, 30, 40, 50, and 60 s]) of both catheters were analyzed and compared. Results: One hundred eighty RF lesions were created in protocol 1 and 300 in protocol 2. The lesion formation, impedance changes, and steam pops characteristics were similar between the two catheters. Higher CFs were related to higher incidences of steam pops. A nonlinear, time‐dependent increase in the lesion depth and diameter was observed for all power and CF settings, and linear, positive correlations between the RF delivery time and lesion volume were observed for all power settings. Compared with 40 W, a 50 W ablation created greater lesions. Longer durations with higher CF settings had a higher steam pop incidence. Conclusions: The lesion formation and incidence of steam pops with TactiFlex SE and FlexAbility SE were similar. A 40 or 50 W ablation withAbstract: Background: Recently, a novel contact force (CF) sensing catheter with mesh‐shaped irrigation tip (TactiFlex SE, Abbott) was invented and is expected to be useful for safe and effective radiofrequency ablation. However, this catheter's detailed characteristics of the lesion formation are unknown. Methods: With an in vitro model, TactiFlex SE and its predecessor, FlexAbility SE, were used. A cross‐sectional analysis of 60 s lesions (combination of various energy power settings [30, 40, and 50 W], and CFs [10, 30, and 50 g]) and longitudinal analysis (combination of various powers [40 or 50 W], CFs [10, 30, and 50 g] and ablation times [10, 20, 30, 40, 50, and 60 s]) of both catheters were analyzed and compared. Results: One hundred eighty RF lesions were created in protocol 1 and 300 in protocol 2. The lesion formation, impedance changes, and steam pops characteristics were similar between the two catheters. Higher CFs were related to higher incidences of steam pops. A nonlinear, time‐dependent increase in the lesion depth and diameter was observed for all power and CF settings, and linear, positive correlations between the RF delivery time and lesion volume were observed for all power settings. Compared with 40 W, a 50 W ablation created greater lesions. Longer durations with higher CF settings had a higher steam pop incidence. Conclusions: The lesion formation and incidence of steam pops with TactiFlex SE and FlexAbility SE were similar. A 40 or 50 W ablation with careful CF control not to exceed 30 g in addition to monitoring impedance drops was required to safely create transmural lesions. Abstract : Lesion formation using a TactiFlex catheter. … (more)
- Is Part Of:
- Journal of arrhythmia. Volume 39:Issue 2(2023)
- Journal:
- Journal of arrhythmia
- Issue:
- Volume 39:Issue 2(2023)
- Issue Display:
- Volume 39, Issue 2 (2023)
- Year:
- 2023
- Volume:
- 39
- Issue:
- 2
- Issue Sort Value:
- 2023-0039-0002-0000
- Page Start:
- 166
- Page End:
- 174
- Publication Date:
- 2023-02-27
- Subjects:
- contact force -- ex vivo experimental model -- lesion size -- radiofrequency ablation -- TactiFlex catheter
Arrhythmia -- Periodicals
Cardiac pacing -- Periodicals
Arrhythmias, Cardiac
Arrhythmia
Cardiac pacing
Periodicals
Electronic journals
Periodicals
616.128 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1883-2148/issues ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/joa3.12835 ↗
- Languages:
- English
- ISSNs:
- 1880-4276
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26900.xml