Anterior vs. posterior position of dispersive patch during radiofrequency catheter ablation: insights from in silico modelling. Issue 3 (13th January 2023)
- Record Type:
- Journal Article
- Title:
- Anterior vs. posterior position of dispersive patch during radiofrequency catheter ablation: insights from in silico modelling. Issue 3 (13th January 2023)
- Main Title:
- Anterior vs. posterior position of dispersive patch during radiofrequency catheter ablation: insights from in silico modelling
- Authors:
- Irastorza, Ramiro M
Maher, Timothy
Barkagan, Michael
Liubasuskas, Rokas
Berjano, Enrique
d'Avila, Andre - Abstract:
- Abstract: Aims: To test the hypothesis that the dispersive patch (DP) location does not significantly affect the current distribution around the catheter tip during radiofrequency catheter ablation (RFCA) but may affect lesions size through differences in impedance due to factors far from the catheter tip. Methods: An in silico model of RFCA in the posterior left atrium and anterior right ventricle was created using anatomic measurements from patient thoracic computed tomography scans and tested the effect of anterior vs. posterior DP locations on baseline impedance, myocardial power delivery, radiofrequency current path, and predicted lesion size. Results: For posterior left atrium ablation, the baseline impedance, total current delivered, current distribution, and proportion of power delivered to the myocardium were all similar with both anterior and posterior DP locations, resulting in similar RFCA lesion sizes (< 0.2 mm difference). For anterior right ventricular (RV) ablation, an anterior DP location resulted in slightly higher proportion of power delivered to the myocardium and lower baseline impedance leading to slightly larger RFCA lesions (0.6 mm deeper and 0.8 mm wider). Conclusions: An anterior vs. posterior DP location will not meaningfully affect RFCA for posterior left atrial ablation, and the slightly larger lesions predicted with anterior DP location for anterior RV ablation are of unclear clinical significance. Graphical Abstract: Graphical Abstract InAbstract: Aims: To test the hypothesis that the dispersive patch (DP) location does not significantly affect the current distribution around the catheter tip during radiofrequency catheter ablation (RFCA) but may affect lesions size through differences in impedance due to factors far from the catheter tip. Methods: An in silico model of RFCA in the posterior left atrium and anterior right ventricle was created using anatomic measurements from patient thoracic computed tomography scans and tested the effect of anterior vs. posterior DP locations on baseline impedance, myocardial power delivery, radiofrequency current path, and predicted lesion size. Results: For posterior left atrium ablation, the baseline impedance, total current delivered, current distribution, and proportion of power delivered to the myocardium were all similar with both anterior and posterior DP locations, resulting in similar RFCA lesion sizes (< 0.2 mm difference). For anterior right ventricular (RV) ablation, an anterior DP location resulted in slightly higher proportion of power delivered to the myocardium and lower baseline impedance leading to slightly larger RFCA lesions (0.6 mm deeper and 0.8 mm wider). Conclusions: An anterior vs. posterior DP location will not meaningfully affect RFCA for posterior left atrial ablation, and the slightly larger lesions predicted with anterior DP location for anterior RV ablation are of unclear clinical significance. Graphical Abstract: Graphical Abstract In silico model of radiofrequency ablation— Human thoracic computed tomography measurements allow the creation of realistic computer simulations of current delivery during ablation of different locations in the heart. Current distribution can be modelled to show differences in ablation with an anterior vs. posterior dispersive patch location. … (more)
- Is Part Of:
- Europace. Volume 25:Issue 3(2023)
- Journal:
- Europace
- Issue:
- Volume 25:Issue 3(2023)
- Issue Display:
- Volume 25, Issue 3 (2023)
- Year:
- 2023
- Volume:
- 25
- Issue:
- 3
- Issue Sort Value:
- 2023-0025-0003-0000
- Page Start:
- 1135
- Page End:
- 1143
- Publication Date:
- 2023-01-13
- Subjects:
- Computer modelling -- Dispersive patch -- In silico model -- Radiofrequency ablation
Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euac277 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26802.xml