Propensity matched analysis of vascular complications using integrated or expandable sheaths for TAVR. Issue 5 (23rd January 2022)
- Record Type:
- Journal Article
- Title:
- Propensity matched analysis of vascular complications using integrated or expandable sheaths for TAVR. Issue 5 (23rd January 2022)
- Main Title:
- Propensity matched analysis of vascular complications using integrated or expandable sheaths for TAVR
- Authors:
- Ruge, Hendrik
Burri, Melchior
Erlebach, Magdalena
Lange, Ruediger - Abstract:
- Abstract: Objectives: Vascular access site complications increase morbidity and mortality in transcatheter aortic valve replacement (TAVR). Background: Medtronic's EnVeo PRO® low‐profile sheath concept and Edwards' expandable eSheath® aim to lower vascular trauma and access site complications. This study aims to compare Valve Academic Research Consortium (VARC)‐3 defined access‐related vascular complications using the two different transcatheter heart valve (THV) delivery concepts. Methods: We performed a retrospective, propensity‐matched study to compare access site vascular complications in 756 consecutive patients who underwent a transfemoral TAVR using a Medtronic Evolut‐R®/Evolut‐PRO® or an Edwards Sapien3®/Sapien3ultra® THV. Results: Propensity score matching resulted in 275 patient pairs. The primary endpoint of major VARC‐3 vascular complication was 7.6% in the Medtronic group and 12.7% in the Edwards group ( p = 0.066). Minor VARC‐3 vascular complications were 9.1% and 8%, respectively ( p = 0.76). VARC‐3 bleeding complications (8.4% vs. 12.7%, p = 0.129) length of hospital stay (7.6 + 5.4 vs. 7.5 + 3.7 days, p = 0.783) and in‐hospital mortality (1.1% vs. 0.4%, p = 0.624) were comparable between both groups. Conclusions: In a propensity‐matched TAVR population, patients treated with the integrated sheath showed a trend towards fewer major vascular complications than patients treated with an expandable sheath, however, the difference was not statisticallyAbstract: Objectives: Vascular access site complications increase morbidity and mortality in transcatheter aortic valve replacement (TAVR). Background: Medtronic's EnVeo PRO® low‐profile sheath concept and Edwards' expandable eSheath® aim to lower vascular trauma and access site complications. This study aims to compare Valve Academic Research Consortium (VARC)‐3 defined access‐related vascular complications using the two different transcatheter heart valve (THV) delivery concepts. Methods: We performed a retrospective, propensity‐matched study to compare access site vascular complications in 756 consecutive patients who underwent a transfemoral TAVR using a Medtronic Evolut‐R®/Evolut‐PRO® or an Edwards Sapien3®/Sapien3ultra® THV. Results: Propensity score matching resulted in 275 patient pairs. The primary endpoint of major VARC‐3 vascular complication was 7.6% in the Medtronic group and 12.7% in the Edwards group ( p = 0.066). Minor VARC‐3 vascular complications were 9.1% and 8%, respectively ( p = 0.76). VARC‐3 bleeding complications (8.4% vs. 12.7%, p = 0.129) length of hospital stay (7.6 + 5.4 vs. 7.5 + 3.7 days, p = 0.783) and in‐hospital mortality (1.1% vs. 0.4%, p = 0.624) were comparable between both groups. Conclusions: In a propensity‐matched TAVR population, patients treated with the integrated sheath showed a trend towards fewer major vascular complications than patients treated with an expandable sheath, however, the difference was not statistically significant. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 99:Issue 5(2022)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 99:Issue 5(2022)
- Issue Display:
- Volume 99, Issue 5 (2022)
- Year:
- 2022
- Volume:
- 99
- Issue:
- 5
- Issue Sort Value:
- 2022-0099-0005-0000
- Page Start:
- 1611
- Page End:
- 1618
- Publication Date:
- 2022-01-23
- Subjects:
- aortic valve disease -- ascular access -- complicationsv -- percutaneous intervention -- transcatheter valve implantation
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.30088 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 26754.xml