Defects in sphygmomanometers: an important source of error in blood pressure recording. Issue 6014 (10th April 1976)
- Record Type:
- Journal Article
- Title:
- Defects in sphygmomanometers: an important source of error in blood pressure recording. Issue 6014 (10th April 1976)
- Main Title:
- Defects in sphygmomanometers: an important source of error in blood pressure recording.
- Authors:
- Conceiçao, S
Ward, M K
Kerr, D N - Abstract:
- Abstract : Almost half the sphygmomanometers in a teaching hospital group had defects in the control valve which interfered with accurate blood-pressure reading. Ward staff should be taught to check sphygmomanometers regularly and replace control valves; time consumption and cost are low. The cuffs in general use in these hospitals, and of the standard size sold in Britain, had rubber bags which did not encircle the arm of more than half the patients on whom they were used. This deficiency causes over-reading of blood pressure in obese people. The size of error is uncertain but it should be avoided by adopting a cuff with a longer bag.
- Is Part Of:
- BMJ. Volume 1:Issue 6014(1976)
- Journal:
- BMJ
- Issue:
- Volume 1:Issue 6014(1976)
- Issue Display:
- Volume 1, Issue 6014 (1976)
- Year:
- 1976
- Volume:
- 1
- Issue:
- 6014
- Issue Sort Value:
- 1976-0001-6014-0000
- Page Start:
- 886
- Page End:
- 888
- Publication Date:
- 1976-04-10
- Subjects:
- Medicine -- Periodicals
Medicine -- Periodicals
Medicine
Periodicals
610 - Journal URLs:
- http://www.bmj.com/archive ↗
http://www.jstor.org/journals/09598138.html ↗
http://www.ncbi.nlm.nih.gov/pmc/journals/3/ ↗
http://www.bmj.com/bmj/ ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/bmj.1.6014.886 ↗
- Languages:
- English
- ISSNs:
- 0007-1447
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26497.xml