Estimation of Goldmann applanation tonometer intraocular pressure (IOP) from scleral Schiotz IOP values in eyes with type-1 keratoprostheses. Issue 4 (8th December 2020)
- Record Type:
- Journal Article
- Title:
- Estimation of Goldmann applanation tonometer intraocular pressure (IOP) from scleral Schiotz IOP values in eyes with type-1 keratoprostheses. Issue 4 (8th December 2020)
- Main Title:
- Estimation of Goldmann applanation tonometer intraocular pressure (IOP) from scleral Schiotz IOP values in eyes with type-1 keratoprostheses
- Authors:
- Senthil, Sirisha
Chary, Raghava
Ali, Mohammed Hasnat
Choudhari, Nikhil S
Garudadri, Chandra Sekhar - Abstract:
- Abstract : Purpose: To validate estimation of Goldmann applanation tonometer (GAT) intraocular pressure (IOP) from scleral Schiotz IOP measurements using a regression model in normal eyes and eyes with type-1 keratoprostheses. Methods: In this prospective cross-sectional study, cohort-1 had 253 normal anterior segment eyes, and cohort-2 had 100 eyes with type-1 keratoprostheses. Scleral Schiotz IOP measurements were used (in a non-linear model) to predict GAT IOP values for these eyes. Accuracy of predicted GAT IOP values was assessed using actual GAT IOP values for normal eyes, while for type-1 keratoprosthetic eyes, finger tension (FT) IOP assessments by an experienced glaucoma specialist were used. Primary outcome was agreement between FT IOP (assessed by an experienced glaucoma specialist) and predicted GAT IOP-derived clusters. Results: The actual values of GAT IOP measurements in normal eyes (n=253; mean age ±SD, 51.35±15.56 years) ranged between 6 mm Hg and 62 mm Hg (mean=22±10.05 mm Hg). Estimated and actual GAT IOP values for normal eyes were very similar (mean difference=0.05 mm Hg with limits of agreement: −5.39 to 5.5 by Bland-Altman plot). Of the 100 eyes with type-1 keratoprostheses, 68 were classified as having digitally normal IOP, 28 as borderline and 4 as high. The agreement between classification by FT assessment and model-predicted GAT IOP values was substantial (Kappa=0.81, 95% CI 0.69 to 0.93). The accuracy of the model in assessing IOP was found to beAbstract : Purpose: To validate estimation of Goldmann applanation tonometer (GAT) intraocular pressure (IOP) from scleral Schiotz IOP measurements using a regression model in normal eyes and eyes with type-1 keratoprostheses. Methods: In this prospective cross-sectional study, cohort-1 had 253 normal anterior segment eyes, and cohort-2 had 100 eyes with type-1 keratoprostheses. Scleral Schiotz IOP measurements were used (in a non-linear model) to predict GAT IOP values for these eyes. Accuracy of predicted GAT IOP values was assessed using actual GAT IOP values for normal eyes, while for type-1 keratoprosthetic eyes, finger tension (FT) IOP assessments by an experienced glaucoma specialist were used. Primary outcome was agreement between FT IOP (assessed by an experienced glaucoma specialist) and predicted GAT IOP-derived clusters. Results: The actual values of GAT IOP measurements in normal eyes (n=253; mean age ±SD, 51.35±15.56 years) ranged between 6 mm Hg and 62 mm Hg (mean=22±10.05 mm Hg). Estimated and actual GAT IOP values for normal eyes were very similar (mean difference=0.05 mm Hg with limits of agreement: −5.39 to 5.5 by Bland-Altman plot). Of the 100 eyes with type-1 keratoprostheses, 68 were classified as having digitally normal IOP, 28 as borderline and 4 as high. The agreement between classification by FT assessment and model-predicted GAT IOP values was substantial (Kappa=0.81, 95% CI 0.69 to 0.93). The accuracy of the model in assessing IOP was found to be 91% (95% CI 0.84 to 0.96). Conclusion: Scleral Schiotz IOP values along with our predictive model can be an alternative objective method to FT IOP in assessing IOP in eyes with type-1 keratoprostheses. … (more)
- Is Part Of:
- British journal of ophthalmology. Volume 106:Issue 4(2022)
- Journal:
- British journal of ophthalmology
- Issue:
- Volume 106:Issue 4(2022)
- Issue Display:
- Volume 106, Issue 4 (2022)
- Year:
- 2022
- Volume:
- 106
- Issue:
- 4
- Issue Sort Value:
- 2022-0106-0004-0000
- Page Start:
- 480
- Page End:
- 484
- Publication Date:
- 2020-12-08
- Subjects:
- cornea -- glaucoma -- intraocular pressure -- ocular surface -- prosthesis
Ophthalmology -- Periodicals
617.7 - Journal URLs:
- http://bjo.bmj.com/ ↗
http://bjo.bmjjournals.com/ ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/bjophthalmol-2020-318121 ↗
- Languages:
- English
- ISSNs:
- 0007-1161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26362.xml