A retrospective study of 12, 538 internal conical connection implants focused on the long‐term integrity of implant–abutment complexes. Issue 4 (16th February 2022)
- Record Type:
- Journal Article
- Title:
- A retrospective study of 12, 538 internal conical connection implants focused on the long‐term integrity of implant–abutment complexes. Issue 4 (16th February 2022)
- Main Title:
- A retrospective study of 12, 538 internal conical connection implants focused on the long‐term integrity of implant–abutment complexes
- Authors:
- Jin, Xiaoting
Guan, Ye
Ren, Jie
Zhao, Yanwei
Wang, Xiaoxiang
He, Fuming - Abstract:
- Abstract: Objective: To evaluate the long‐term integrity of implant–abutment complexes in implant systems with two internal conical angles. Material and Methods: 12, 538 bone‐level implants of two systems placed between January 2012 and December 2018 were retrospectively analyzed. Cumulative abutment/implant fracture rates in systems with larger (LA, 7.5°) and smaller (SA, 5.7°) internal conical angles were estimated using the Kaplan–Meier analysis and compared between groups. The association between implant systems and jammed abutment retrievability was evaluated by multivariable generalized estimating equation logistic regression modeling. Results: For LA, the 8‐year cumulative incident rate was 0.10% (95% confidence interval (CI): 0–0.24%) for implant fracture and 0.26% (95% CI: 0.11%–0.41%) for abutment fracture, demonstrating a significant difference in gender ( p = .03), implant diameter ( p = .01), jaw ( p = .006), and antagonist tooth ( p < .001). For SA, the 8‐year cumulative incident rate was 0.38% (95% CI: 0–0.79%) for implant fracture and 2.62% (95% CI: 0.05%–5.13%) for abutment fracture, which was influenced by implant diameter ( p < .001) and site ( p = .03). The cumulative implant/abutment fracture rate was lower for LA implants, particularly for LA implant‐supported single crowns (SCs) ( p < .05). The abutment‐retrieval success rate was 92.9% for LA and 57.1% for SA ( p = .055). Conclusion: LA implants exhibited a lower incidence of fracture inAbstract: Objective: To evaluate the long‐term integrity of implant–abutment complexes in implant systems with two internal conical angles. Material and Methods: 12, 538 bone‐level implants of two systems placed between January 2012 and December 2018 were retrospectively analyzed. Cumulative abutment/implant fracture rates in systems with larger (LA, 7.5°) and smaller (SA, 5.7°) internal conical angles were estimated using the Kaplan–Meier analysis and compared between groups. The association between implant systems and jammed abutment retrievability was evaluated by multivariable generalized estimating equation logistic regression modeling. Results: For LA, the 8‐year cumulative incident rate was 0.10% (95% confidence interval (CI): 0–0.24%) for implant fracture and 0.26% (95% CI: 0.11%–0.41%) for abutment fracture, demonstrating a significant difference in gender ( p = .03), implant diameter ( p = .01), jaw ( p = .006), and antagonist tooth ( p < .001). For SA, the 8‐year cumulative incident rate was 0.38% (95% CI: 0–0.79%) for implant fracture and 2.62% (95% CI: 0.05%–5.13%) for abutment fracture, which was influenced by implant diameter ( p < .001) and site ( p = .03). The cumulative implant/abutment fracture rate was lower for LA implants, particularly for LA implant‐supported single crowns (SCs) ( p < .05). The abutment‐retrieval success rate was 92.9% for LA and 57.1% for SA ( p = .055). Conclusion: LA implants exhibited a lower incidence of fracture in abutment–implant complexes and a relatively higher retrievability success rate for jammed abutments. … (more)
- Is Part Of:
- Clinical oral implants research. Volume 33:Issue 4(2022)
- Journal:
- Clinical oral implants research
- Issue:
- Volume 33:Issue 4(2022)
- Issue Display:
- Volume 33, Issue 4 (2022)
- Year:
- 2022
- Volume:
- 33
- Issue:
- 4
- Issue Sort Value:
- 2022-0033-0004-0000
- Page Start:
- 377
- Page End:
- 390
- Publication Date:
- 2022-02-16
- Subjects:
- abutment fracture -- implant fracture -- internal conical connection -- metal fatigue
Dental implants -- Research -- Periodicals
617.69 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/clr.13898 ↗
- Languages:
- English
- ISSNs:
- 0905-7161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.318000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 26358.xml