Feasibility of operating a millimeter‐scale graphite calorimeter for absolute dosimetry of small‐field photon beams in the clinic. Issue 11 (5th October 2021)
- Record Type:
- Journal Article
- Title:
- Feasibility of operating a millimeter‐scale graphite calorimeter for absolute dosimetry of small‐field photon beams in the clinic. Issue 11 (5th October 2021)
- Main Title:
- Feasibility of operating a millimeter‐scale graphite calorimeter for absolute dosimetry of small‐field photon beams in the clinic
- Authors:
- Côté, Benjamin
Keszti, Federico
Bancheri, Julien
Sarfehnia, Arman
Seuntjens, Jan
Renaud, James - Abstract:
- Abstract: Purpose: To characterize and build a cylindrically layered graphite calorimeter the size of a thimble ionization chamber for absolute dosimetry of small fields. This detector has been designed in a familiar probe format to facilitate integration into the clinical workflow. The feasibility of operating this absorbed dose calorimeter in quasi‐adiabatic mode is assessed for high‐energy accelerator‐based photon beams. Methods: This detector, herein referred to as Aerrow MK7, is a miniaturized version of a previously validated aerogel‐insulated graphite calorimeter known as Aerrow. The new model was designed and developed using numerical methods. Medium conversion factors from graphite to water, small‐field output correction factors, and layer perturbation factors for this dosimeter were calculated using the EGSnrc Monte Carlo code system. A range of commercially available aerogel densities were studied for the insulating layers, and an optimal density was selected by minimizing the small‐field output correction factors. Heat exchange within the detector was simulated using a five‐body compartmental heat transfer model. In quasi‐adiabatic mode, the sensitive volume (a 3 mm diameter cylindrical graphite core) experiences a temperature rise during irradiation on the order of 1.3 mK·Gy −1 . The absorbed dose is obtained by calculating the product of this temperature rise with the specific heat capacity of the graphite. The detector was irradiated with 6 MV ( % dd ( 10 ) xAbstract: Purpose: To characterize and build a cylindrically layered graphite calorimeter the size of a thimble ionization chamber for absolute dosimetry of small fields. This detector has been designed in a familiar probe format to facilitate integration into the clinical workflow. The feasibility of operating this absorbed dose calorimeter in quasi‐adiabatic mode is assessed for high‐energy accelerator‐based photon beams. Methods: This detector, herein referred to as Aerrow MK7, is a miniaturized version of a previously validated aerogel‐insulated graphite calorimeter known as Aerrow. The new model was designed and developed using numerical methods. Medium conversion factors from graphite to water, small‐field output correction factors, and layer perturbation factors for this dosimeter were calculated using the EGSnrc Monte Carlo code system. A range of commercially available aerogel densities were studied for the insulating layers, and an optimal density was selected by minimizing the small‐field output correction factors. Heat exchange within the detector was simulated using a five‐body compartmental heat transfer model. In quasi‐adiabatic mode, the sensitive volume (a 3 mm diameter cylindrical graphite core) experiences a temperature rise during irradiation on the order of 1.3 mK·Gy −1 . The absorbed dose is obtained by calculating the product of this temperature rise with the specific heat capacity of the graphite. The detector was irradiated with 6 MV ( % dd ( 10 ) x = 63.5%) and 10 MV ( % dd ( 10 ) x = 71.1%) flattening filter‐free (FFF) photon beams for two field sizes, characterized by S clin dimensions of 2.16 and 11.0 cm. The dose readings were compared against a calibrated Exradin A1SL ionization chamber. All dose values are reported at d max in water. Results: The field output correction factors for this dosimeter design were computed for field sizes ranging from S clin = 0.54 to 11.0 cm. For all aerogel densities studied, these correction factors did not exceed 1.5%. The relative dose difference between the two dosimeters ranged between 0.3% and 0.7% for all beams and field sizes. The smallest field size experimentally investigated, S clin = 2.16 cm, which was irradiated with the 10 MV FFF beam, produced readings of 84.4 cGy (±1.3%) in the calorimeter and 84.5 cGy (±1.3%) in the ionization chamber. Conclusion: The median relative difference in absorbed dose values between a calibrated A1SL ionization chamber and the proposed novel graphite calorimeter was 0.6%. This preliminary experimental validation demonstrates that Aerrow MK7 is capable of accurate and reproducible absorbed dose measurements in quasi‐adiabatic mode. … (more)
- Is Part Of:
- Medical physics. Volume 48:Issue 11(2021)
- Journal:
- Medical physics
- Issue:
- Volume 48:Issue 11(2021)
- Issue Display:
- Volume 48, Issue 11 (2021)
- Year:
- 2021
- Volume:
- 48
- Issue:
- 11
- Issue Sort Value:
- 2021-0048-0011-0000
- Page Start:
- 7476
- Page End:
- 7492
- Publication Date:
- 2021-10-05
- Subjects:
- calorimetry -- heat transfer modeling -- small‐field dosimetry
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.15244 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26352.xml