The Interfield Strength Agreement of Left Ventricular Strain Measurements at 1.5 T and 3 T Using Cardiac MRI Feature Tracking. Issue 4 (29th June 2022)
- Record Type:
- Journal Article
- Title:
- The Interfield Strength Agreement of Left Ventricular Strain Measurements at 1.5 T and 3 T Using Cardiac MRI Feature Tracking. Issue 4 (29th June 2022)
- Main Title:
- The Interfield Strength Agreement of Left Ventricular Strain Measurements at 1.5 T and 3 T Using Cardiac MRI Feature Tracking
- Authors:
- Ayton, Sarah L.
Alfuhied, Aseel
Gulsin, Gaurav S.
Parke, Kelly S.
Wormleighton, Joanne V.
Arnold, J. Ranjit
Moss, Alastair J.
Singh, Anvesha
Xue, Hui
Kellman, Peter
Graham‐Brown, Matthew P. M.
McCann, Gerry P. - Abstract:
- Abstract : Background: Left ventricular (LV) strain measurements can be derived using cardiac MRI from routinely acquired balanced steady‐state free precession (bSSFP) cine images. Purpose: To compare the interfield strength agreement of global systolic strain, peak strain rates and artificial intelligence (AI) landmark‐based global longitudinal shortening at 1.5 T and 3 T. Study Type: Prospective. Subjects: A total of 22 healthy individuals (mean age 36 ± 12 years; 45% male) completed two cardiac MRI scans at 1.5 T and 3 T in a randomized order within 30 minutes. Field Strength/Sequence: bSSFP cine images at 1.5 T and 3 T. Assessment: Two software packages, Tissue Tracking (cvi42, Circle Cardiovascular Imaging) and QStrain (Medis Suite, Medis Medical Imaging Systems), were used to derive LV global systolic strain in the longitudinal, circumferential and radial directions and peak (systolic, early diastolic, and late diastolic) strain rates. Global longitudinal shortening and mitral annular plane systolic excursion (MAPSE) were measured using an AI deep neural network model. Statistical Tests: Comparisons between field strengths were performed using Wilcoxon signed‐rank test ( P value < 0.05 considered statistically significant). Agreement was determined using intraclass correlation coefficients (ICCs) and Bland–Altman plots. Results: Minimal bias was seen in all strain and strain rate measurements between field strengths. Using Tissue Tracking, strain and strain rate valuesAbstract : Background: Left ventricular (LV) strain measurements can be derived using cardiac MRI from routinely acquired balanced steady‐state free precession (bSSFP) cine images. Purpose: To compare the interfield strength agreement of global systolic strain, peak strain rates and artificial intelligence (AI) landmark‐based global longitudinal shortening at 1.5 T and 3 T. Study Type: Prospective. Subjects: A total of 22 healthy individuals (mean age 36 ± 12 years; 45% male) completed two cardiac MRI scans at 1.5 T and 3 T in a randomized order within 30 minutes. Field Strength/Sequence: bSSFP cine images at 1.5 T and 3 T. Assessment: Two software packages, Tissue Tracking (cvi42, Circle Cardiovascular Imaging) and QStrain (Medis Suite, Medis Medical Imaging Systems), were used to derive LV global systolic strain in the longitudinal, circumferential and radial directions and peak (systolic, early diastolic, and late diastolic) strain rates. Global longitudinal shortening and mitral annular plane systolic excursion (MAPSE) were measured using an AI deep neural network model. Statistical Tests: Comparisons between field strengths were performed using Wilcoxon signed‐rank test ( P value < 0.05 considered statistically significant). Agreement was determined using intraclass correlation coefficients (ICCs) and Bland–Altman plots. Results: Minimal bias was seen in all strain and strain rate measurements between field strengths. Using Tissue Tracking, strain and strain rate values derived from long‐axis images showed poor to fair agreement (ICC range 0.39–0.71), whereas global longitudinal shortening and MAPSE showed good agreement (ICC = 0.81 and 0.80, respectively). Measures derived from short‐axis images showed good to excellent agreement (ICC range 0.78–0.91). Similar results for the agreement of strain and strain rate measurements were observed with QStrain. Conclusion: The interfield strength agreement of short‐axis derived LV strain and strain rate measurements at 1.5 T and 3 T was better than those derived from long‐axis images; however, the agreement of global longitudinal shortening and MAPSE was good. Evidence Level: 2 Technical Efficacy: Stage 2 … (more)
- Is Part Of:
- Journal of magnetic resonance imaging. Volume 57:Issue 4(2023)
- Journal:
- Journal of magnetic resonance imaging
- Issue:
- Volume 57:Issue 4(2023)
- Issue Display:
- Volume 57, Issue 4 (2023)
- Year:
- 2023
- Volume:
- 57
- Issue:
- 4
- Issue Sort Value:
- 2023-0057-0004-0000
- Page Start:
- 1250
- Page End:
- 1261
- Publication Date:
- 2022-06-29
- Subjects:
- cardiac MRI -- field strength -- left ventricular strain -- myocardial deformation analysis
Magnetic resonance imaging -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2586 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jmri.28328 ↗
- Languages:
- English
- ISSNs:
- 1053-1807
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5010.791000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26302.xml