Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report. Issue 9 (21st September 2021)
- Record Type:
- Journal Article
- Title:
- Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report. Issue 9 (21st September 2021)
- Main Title:
- Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report
- Authors:
- Patel, Bhavisha
Monkhouse, Christopher
Manisty, Charlotte
Papageorgiou, Nikos - Editors:
- Schönbauer, Robert
Green, Peregrine
Arcari, Luca
Kurdi, Hibba
Thomson, Ross - Abstract:
- Abstract: Background: Magnetic resonance (MR) imaging (MRI) for patients with implantable cardiac devices is becoming more routine, with the development of MR conditional devices allowing more patients access to the imaging they need. However, for this to be performed safely, strict protocols must be followed necessitating close collaboration between cardiology and radiology departments. We present a case where mandatory device re-programming of a cardiac resynchronization therapy defibrillator device into MRI mode was not performed pre-scan leading to temporary device dysfunction with no clinical consequences. Case summary: A 72-year-old man presented to a device clinic for a routine device interrogation. An atrial tachycardia response episode was recorded at the same time as the patient reported having undergone an MRI scan at a local centre. The electrogram demonstrated temporary right ventricular loss of capture with standard output programming, and a short episode of oversensing on the atrial and ventricular channel which was not sustained for long enough to meet tachycardia detection. Discussion: We demonstrate two potential electrophysiological effects of MRI on pacemakers, where the device had not been appropriately re-programmed pre-procedure. This illustrates that whilst MRI in patients with implantable cardiac devices is safe, strict protocols must be followed requiring robust multidisciplinary communication.
- Is Part Of:
- European heart journal. Volume 5:Issue 9(2021)
- Journal:
- European heart journal
- Issue:
- Volume 5:Issue 9(2021)
- Issue Display:
- Volume 5, Issue 9 (2021)
- Year:
- 2021
- Volume:
- 5
- Issue:
- 9
- Issue Sort Value:
- 2021-0005-0009-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-09-21
- Subjects:
- MRI -- Pacemakers -- Implantable cardioverter-defibrillator -- Cardiac resynchronization defibrillator -- Loss of capture -- Programming -- MRI safety -- Case report
Cardiology -- Periodicals
Cardiology -- Case studies -- Periodicals
Heart -- Diseases -- Periodicals
Heart -- Diseases -- Case studies -- Periodicals
616.12 - Journal URLs:
- http://www.oxfordjournals.org/ ↗
https://academic.oup.com/ehjcr/ ↗ - DOI:
- 10.1093/ehjcr/ytab198 ↗
- Languages:
- English
- ISSNs:
- 2514-2119
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26305.xml