Failure of knee arthroplasty secondary to inadequate technique in cephalomedullary nailing of a subtrochanteric hip fracture. (3rd April 2012)
- Record Type:
- Journal Article
- Title:
- Failure of knee arthroplasty secondary to inadequate technique in cephalomedullary nailing of a subtrochanteric hip fracture. (3rd April 2012)
- Main Title:
- Failure of knee arthroplasty secondary to inadequate technique in cephalomedullary nailing of a subtrochanteric hip fracture
- Authors:
- Clement, Rhys Gareth Ellis
Burnett, Richard - Abstract:
- Abstract : A case of total knee replacement loosening caused by a failed cephalomedullary nail is presented. The nail was inserted for the treatment of an unstable subtrochanteric femoral fracture. The distal locking screw of the nail subsequently fractured allowing the leg to shorten and causing the nail to displace the femoral component of the knee replacement leading to a valgus alignment and deformity of the knee. In order to revise the knee the long γ nail had to be replaced by a shorter nail to allow a femoral component with a long stem to be used. This case illustrates the importance of using shorter nails in patients with prosthetic knees.
- Is Part Of:
- BMJ case reports. Volume 2012
- Journal:
- BMJ case reports
- Issue:
- Volume 2012
- Issue Display:
- Volume 2012 (2012)
- Year:
- 2012
- Volume:
- 2012
- Issue Sort Value:
- 2012-2012-0000-0000
- Page Start:
- Page End:
- Publication Date:
- 2012-04-03
- Subjects:
- Medicine -- Case studies -- Periodicals
610.5 - Journal URLs:
- http://www.bmj.com/archive ↗
http://casereports.bmj.com/ ↗ - DOI:
- 10.1136/bcr.12.2011.5430 ↗
- Languages:
- English
- ISSNs:
- 1757-790X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26222.xml