Neutron‐gamma dosimetry for BNCT using field oxide transistors with gadolinium oxide as neutron converter layer. Issue 2 (23rd December 2021)
- Record Type:
- Journal Article
- Title:
- Neutron‐gamma dosimetry for BNCT using field oxide transistors with gadolinium oxide as neutron converter layer. Issue 2 (23rd December 2021)
- Main Title:
- Neutron‐gamma dosimetry for BNCT using field oxide transistors with gadolinium oxide as neutron converter layer
- Authors:
- Gimenez, Melisa Lucía
Lipovetzky, José
Alcalde Bessia, Fabricio
Longhino, Juan Manuel
Tartaglione, Aureliano
Garcia‐Inza, Mariano Andrés
Blostein, Juan Jerónimo
Carbonetto, Sebastián
Gómez Berisso, Mariano
Pérez, Martín
Sidelnik, Iván
Redin, Eduardo Gabriel
Faigón, Adrián - Abstract:
- Abstract: Purpose: A new type of electronic dosimeter is presented, capable of discerning between the doses of gamma photons and neutrons in a mixed beam as found in boron neutron capture therapy (BNCT). We introduce a real‐time dosimeter based on a thick gate field oxide field effect transistor (FOXFET) covered with a neutron converter layer containing gadolinium. Methods: To sensitize the FOXFET dosimeter to neutron fluxes, a converter layer containing gadolinium oxide particles embedded in photoresines was deposited over the sensor surface. Mixed neutron‐gamma field configurations with different neutron energy spectra were used to assess the FOXFET response, considering different thicknesses of the neutron converter layer. Results: The total gamma sensitivity of the devices resulted to be 43 mV/Gy. The responses of sensors with different converter layer thicknesses irradiated with different neutron spectra are simulated using GEANT4 code. The response to photons is not significantly modified with thin conversion layers when used in water medium. Conclusions: A real‐time dosimeter comprising a pair of FOXFET sensors—only one of them with a gadolinium neutron converter layer—allows the simultaneous measurement of gamma dose and neutron flux during BNCT irradiations.
- Is Part Of:
- Medical physics. Volume 49:Issue 2(2022)
- Journal:
- Medical physics
- Issue:
- Volume 49:Issue 2(2022)
- Issue Display:
- Volume 49, Issue 2 (2022)
- Year:
- 2022
- Volume:
- 49
- Issue:
- 2
- Issue Sort Value:
- 2022-0049-0002-0000
- Page Start:
- 1276
- Page End:
- 1285
- Publication Date:
- 2021-12-23
- Subjects:
- boron neutron capture therapy (BNCT) -- neutron‐gamma dosimeter -- radiation effects -- solid‐state dosimeters
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.15385 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26185.xml