Usability of magnetic resonance images acquired at a novel low-field 0.55 T scanner for brain radiotherapy treatment planning. (January 2023)
- Record Type:
- Journal Article
- Title:
- Usability of magnetic resonance images acquired at a novel low-field 0.55 T scanner for brain radiotherapy treatment planning. (January 2023)
- Main Title:
- Usability of magnetic resonance images acquired at a novel low-field 0.55 T scanner for brain radiotherapy treatment planning
- Authors:
- Grigo, Johanna
Masitho, Siti
Fautz, Hans-Peter
Voigt, Raphaela
Schonath, Miriam
Oleszczuk, Annika
Uder, Michael
Heiss, Rafael
Fietkau, Rainer
Putz, Florian
Bert, Christoph - Abstract:
- Abstract: Background and Purpose: Low-field magnetic resonance imaging (MRI) may offer specific advantages over high-field MRI, e.g. lower susceptibility-dependent distortions and simpler installation. The study aim was to evaluate if a novel 0.55 T MRI scanner provides sufficient image accuracy and quality for radiotherapy (RT) treatment planning. Material and methods: The geometric accuracy of images acquired at a low-field MRI scanner was evaluated in phantom measurements regarding gradient non-linearity-related distortions. Patient-induced B0-susceptibility changes were investigated via B0-field-mapping in ten volunteers. Patients were positioned in RT-setup using a 3D-printed insert for the head/neck-coil that was tested for sufficient signal-to-noise-ratio (SNR). The suitability of the MRI-system for detection of metastases was evaluated in eleven patients. In comparison to diagnostic images, acquired at ≥1.5 T, three physicians evaluated the detectability of metastases by counting them in low- and high-field-images, respectively. Results: The phantom measurements showed a high imaging fidelity after 3D-distortion-correction with (1.2 ± 0.9) mm geometric distortion in 10 cm radius from isocentre. At the edges remaining distortions were greater than at 1.5 T. The mean susceptibility-induced distortions in the head were (0.05 ± 0.05) mm and maximum 0.69 mm. SNR analysis showed that optimised positioning of RT-patients without signal loss in the head/neck-coil wasAbstract: Background and Purpose: Low-field magnetic resonance imaging (MRI) may offer specific advantages over high-field MRI, e.g. lower susceptibility-dependent distortions and simpler installation. The study aim was to evaluate if a novel 0.55 T MRI scanner provides sufficient image accuracy and quality for radiotherapy (RT) treatment planning. Material and methods: The geometric accuracy of images acquired at a low-field MRI scanner was evaluated in phantom measurements regarding gradient non-linearity-related distortions. Patient-induced B0-susceptibility changes were investigated via B0-field-mapping in ten volunteers. Patients were positioned in RT-setup using a 3D-printed insert for the head/neck-coil that was tested for sufficient signal-to-noise-ratio (SNR). The suitability of the MRI-system for detection of metastases was evaluated in eleven patients. In comparison to diagnostic images, acquired at ≥1.5 T, three physicians evaluated the detectability of metastases by counting them in low- and high-field-images, respectively. Results: The phantom measurements showed a high imaging fidelity after 3D-distortion-correction with (1.2 ± 0.9) mm geometric distortion in 10 cm radius from isocentre. At the edges remaining distortions were greater than at 1.5 T. The mean susceptibility-induced distortions in the head were (0.05 ± 0.05) mm and maximum 0.69 mm. SNR analysis showed that optimised positioning of RT-patients without signal loss in the head/neck-coil was possible with the RT-insert. No significant differences (p = 0.48) in detectability of metastases were found. Conclusion: The 0.55 T MRI system provided sufficiently geometrically accurate and high-resolution images that can be used for RT-planning for brain metastases. Hence, modern low-field MRI may contribute to simply access MRI for RT-planning after further investigations. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 25(2023)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 25(2023)
- Issue Display:
- Volume 25, Issue 2023 (2023)
- Year:
- 2023
- Volume:
- 25
- Issue:
- 2023
- Issue Sort Value:
- 2023-0025-2023-0000
- Page Start:
- Page End:
- Publication Date:
- 2023-01
- Subjects:
- MRI -- Radiotherapy -- Low-Field-MRI
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2023.100412 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26134.xml