Ex vivo blood vessel bioreactor for analysis of the biodegradation of magnesium stent models with and without vessel wall integration. (1st March 2017)
- Record Type:
- Journal Article
- Title:
- Ex vivo blood vessel bioreactor for analysis of the biodegradation of magnesium stent models with and without vessel wall integration. (1st March 2017)
- Main Title:
- Ex vivo blood vessel bioreactor for analysis of the biodegradation of magnesium stent models with and without vessel wall integration
- Authors:
- Wang, Juan
Liu, Lumei
Wu, Yifan
Maitz, Manfred F.
Wang, Zhihong
Koo, Youngmi
Zhao, Ansha
Sankar, Jagannathan
Kong, Deling
Huang, Nan
Yun, Yeoheung - Abstract:
- Graphical abstract: Abstract: Current in vitro models fail in predicting the degradation rate and mode of magnesium (Mg) stents in vivo . To overcome this, the microenvironment of the stent is simulated here in an ex vivo bioreactor with porcine aorta and circulating medium, and compared with standard static in vitro immersion and with in vivo rat aorta models. In ex vivo and in vivo conditions, pure Mg wires were exposed to the aortic lumen and inserted into the aortic wall to mimic early- and long-term implantation, respectively. Results showed that: 1) Degradation rates of Mg were similar for all the fluid diffusion conditions ( in vitro static, aortic wall ex vivo and in vivo ); however, Mg degradation under flow condition (i.e. in the lumen) in vivo was slower than ex vivo ; 2) The corrosion mode in the samples can be mainly described as localized ( in vitro ), mixed localized and uniform ( ex vivo ), and uniform ( in vivo ); 3) Abundant degradation products (MgO/Mg(OH)2 and Ca/P) with gas bubbles accumulated around the localized degradation regions ex vivo, but a uniform and thin degradation product layer was found in vivo . It is concluded that the ex vivo vascular bioreactor provides an improved test setting for magnesium degradation between static immersion and animal experiments and highlights its promising role in bridging degradation behavior and biological response for vascular stent research. Statement of Significance: Magnesium and its alloys are candidatesGraphical abstract: Abstract: Current in vitro models fail in predicting the degradation rate and mode of magnesium (Mg) stents in vivo . To overcome this, the microenvironment of the stent is simulated here in an ex vivo bioreactor with porcine aorta and circulating medium, and compared with standard static in vitro immersion and with in vivo rat aorta models. In ex vivo and in vivo conditions, pure Mg wires were exposed to the aortic lumen and inserted into the aortic wall to mimic early- and long-term implantation, respectively. Results showed that: 1) Degradation rates of Mg were similar for all the fluid diffusion conditions ( in vitro static, aortic wall ex vivo and in vivo ); however, Mg degradation under flow condition (i.e. in the lumen) in vivo was slower than ex vivo ; 2) The corrosion mode in the samples can be mainly described as localized ( in vitro ), mixed localized and uniform ( ex vivo ), and uniform ( in vivo ); 3) Abundant degradation products (MgO/Mg(OH)2 and Ca/P) with gas bubbles accumulated around the localized degradation regions ex vivo, but a uniform and thin degradation product layer was found in vivo . It is concluded that the ex vivo vascular bioreactor provides an improved test setting for magnesium degradation between static immersion and animal experiments and highlights its promising role in bridging degradation behavior and biological response for vascular stent research. Statement of Significance: Magnesium and its alloys are candidates for a new generation of biodegradable stent materials. However, the in vitro degradation of magnesium stents does not match the clinical degradation rates, corrupting the validity of conventional degradation tests. Here we report an ex vivo vascular bioreactor, which allows simulation of the microenvironment with and without blood vessel integration to study the biodegradation of magnesium implants in comparison with standard in vitro test conditions and with in vivo implantations. The bioreactor did simulate the corrosion of an intramural implant very well, but showed too high degradation for non-covered implants. It is concluded that this system is in between static incubation and animal experiments concerning the predictivity of the degradation. … (more)
- Is Part Of:
- Acta biomaterialia. Volume 50(2017)
- Journal:
- Acta biomaterialia
- Issue:
- Volume 50(2017)
- Issue Display:
- Volume 50, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 50
- Issue:
- 2017
- Issue Sort Value:
- 2017-0050-2017-0000
- Page Start:
- 546
- Page End:
- 555
- Publication Date:
- 2017-03-01
- Subjects:
- Magnesium -- Biodegradation -- Vascular bioreactor -- Ex vivo -- In vitro -- In vivo
Biomedical materials -- Periodicals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17427061 ↗
http://www.elsevier.com/wps/find/journaldescription.cws%5Fhome/702994/description ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.actbio.2016.12.039 ↗
- Languages:
- English
- ISSNs:
- 1742-7061
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0602.900500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 26132.xml