Tateshima, Satoshi - Abstract:
- Abstract : Endovascular treatment of broad-necked bifurcation aneurysms remains challenging. Stent-assisted coiling has been successful but requires catheterization of the branches off the parent vessel. We present the case of a patient who failed primary and stent-assisted coiling of a large basilar tip aneurysm because the morphology of the aneurysm precluded successful distal catheterization of the posterior cerebral artery (PCA) branches. Using the PulseRider device, which does not require catheterization of bifurcation branches, we were able to treat the aneurysm successfully.
- Is Part Of:
- BMJ case reports. Volume 2015
- Journal:
- BMJ case reports
- Issue:
- Volume 2015
- Issue Display:
- Volume 2015 (2015)
- Year:
- 2015
- Volume:
- 2015
- Issue Sort Value:
- 2015-2015-0000-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-07-21
- Subjects:
- Aneurysm -- Coil -- Device -- Stent -- Technique
Medicine -- Case studies -- Periodicals
610.5 - Journal URLs:
- http://www.bmj.com/archive ↗
http://casereports.bmj.com/ ↗ - DOI:
- 10.1136/bcr-2015-011836 ↗
- Languages:
- English
- ISSNs:
- 1757-790X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25913.xml