Report of Task Group 201 of the American Association of Physicists in Medicine: Quality management of external beam therapy data transfer. Issue 6 (23rd April 2021)
- Record Type:
- Journal Article
- Title:
- Report of Task Group 201 of the American Association of Physicists in Medicine: Quality management of external beam therapy data transfer. Issue 6 (23rd April 2021)
- Main Title:
- Report of Task Group 201 of the American Association of Physicists in Medicine: Quality management of external beam therapy data transfer
- Authors:
- Siochi, R. Alfredo
Balter, Peter
Bloch, Charles D.
Santanam, Lakshmi
Blodgett, Kurt
Curran, Bruce H.
Engelsman, Martijn
Feng, Wenzheng
Mechalakos, James
Pavord, Daniel
Simon, Thomas
Sutlief, Steven
Zhu, X. Ronald - Abstract:
- Abstract : With the advancement of data‐intensive technologies, such as image‐guided radiation therapy (IGRT) and intensity‐modulated radiation therapy (IMRT), the amount and complexity of data to be transferred between clinical subsystems have increased beyond the reach of manual checking. As a result, unintended treatment deviations (e.g., dose errors) may occur if the treatment system is not closely monitored by a comprehensive data transfer quality management program (QM). This report summarizes the findings and recommendations from the task group (TG) on quality assurance (QA) of external beam treatment data transfer (TG‐201), with the aim to assist medical physicists in designing their own data transfer QM. As a background, a section of this report describes various models of data flow (distributed data repositories and single data base systems) and general data test characteristics (data integrity, interpretation, and consistency). Recommended tests are suggested based on the collective experience of TG‐201 members. These tests are for the acceptance of, commissioning of, and upgrades to subsystems that store and/or modify clinical treatment data. As treatment complexity continues to evolve, we will need to do and know more about ensuring the quality of data transfers. The report concludes with the recommendation to move toward data transfer open standards compatibility and to develop tools that automate data transfer QA.
- Is Part Of:
- Medical physics. Volume 48:Issue 6(2021)
- Journal:
- Medical physics
- Issue:
- Volume 48:Issue 6(2021)
- Issue Display:
- Volume 48, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 48
- Issue:
- 6
- Issue Sort Value:
- 2021-0048-0006-0000
- Page Start:
- e86
- Page End:
- e114
- Publication Date:
- 2021-04-23
- Subjects:
- 29.1 (TH‐ Dataset analysis/biomathematics -- Informatics) -- 31.1 (TH‐ External beam‐ photons -- General (most aspects)) -- data transfer -- information systems -- quality assurance -- quality control -- quality management program -- Taxonomy: 25.1 -- 3(IM/TH‐ Formal quality management tools -- General (most aspects) -- Root Cause Analysis)
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.14868 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25872.xml