The transseptal puncture experience: Safety insights from FDA MAUDE database. Issue 6 (1st May 2021)
- Record Type:
- Journal Article
- Title:
- The transseptal puncture experience: Safety insights from FDA MAUDE database. Issue 6 (1st May 2021)
- Main Title:
- The transseptal puncture experience: Safety insights from FDA MAUDE database
- Authors:
- Morcos, Ramez
Megaly, Michael
Desai, Anand
Alkhouli, Mohamed
Saad, Marwan
Luna, Michael
Garcia, Santiago
Khalili, Houman
Maini, Brijeshwar - Abstract:
- Abstract: Background: There is a paucity of literature on safety and efficacy of various transseptal puncture (TSP) needles. Objectives: To assess the reported mechanisms of failure, complications, and outcomes among the most frequently used transseptal needles in the United States. Methods: We queried the Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database between January 2011 and January 2021 for reports on the most commonly used transseptal needles: NRG (Baylis Medical, Montreal, Canada), and BRK (St. Jude, Saint Paul, MN)]. The primary outcome was the mechanism of failure. Secondary outcomes included clinical consequences of device failure. Results: The final analysis included 306 reports of failure/complication with TSP needles (NRG n = 70, BRK n = 236). The most commonly reported mode of failure was detachment of the needle component (i.e., clip, hub, stopcock, shaft, spring, or needle tip) (14.7% overall; 17.8% BRK; and 4.3% NRG). Among these reports, cardiac perforation was the most common complication (69.9% overall; 69.1% for BRK; and 72.9% for NRG). Pericardiocentesis was the second most commonly reported complication (45.1% overall; 48.3% for BRK; and 34.3% for NRG). The procedure was successfully completed in 33.3% of all cases (36.4% for BRK and 22.9% for NRG), while surgical conversion was needed in (13.4% overall; 14% for BRK and 11.4% for NRG) of the reports. Death occurred in 3.9% of all cases overall (3.4%Abstract: Background: There is a paucity of literature on safety and efficacy of various transseptal puncture (TSP) needles. Objectives: To assess the reported mechanisms of failure, complications, and outcomes among the most frequently used transseptal needles in the United States. Methods: We queried the Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database between January 2011 and January 2021 for reports on the most commonly used transseptal needles: NRG (Baylis Medical, Montreal, Canada), and BRK (St. Jude, Saint Paul, MN)]. The primary outcome was the mechanism of failure. Secondary outcomes included clinical consequences of device failure. Results: The final analysis included 306 reports of failure/complication with TSP needles (NRG n = 70, BRK n = 236). The most commonly reported mode of failure was detachment of the needle component (i.e., clip, hub, stopcock, shaft, spring, or needle tip) (14.7% overall; 17.8% BRK; and 4.3% NRG). Among these reports, cardiac perforation was the most common complication (69.9% overall; 69.1% for BRK; and 72.9% for NRG). Pericardiocentesis was the second most commonly reported complication (45.1% overall; 48.3% for BRK; and 34.3% for NRG). The procedure was successfully completed in 33.3% of all cases (36.4% for BRK and 22.9% for NRG), while surgical conversion was needed in (13.4% overall; 14% for BRK and 11.4% for NRG) of the reports. Death occurred in 3.9% of all cases overall (3.4% for BRK and 5.7% for NRG). Conclusions: Needle detachment was the most common mode of failure, and cardiac perforation was the most common complication reported with TSP needles. Future efforts should focus on innovative TSP needle design, best practice guidelines, including role of imaging guidance, and increased TSP training. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 98:Issue 6(2021)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 98:Issue 6(2021)
- Issue Display:
- Volume 98, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 98
- Issue:
- 6
- Issue Sort Value:
- 2021-0098-0006-0000
- Page Start:
- E855
- Page End:
- E861
- Publication Date:
- 2021-05-01
- Subjects:
- BRK -- MAUDE -- NRG -- transseptal needles -- transseptal puncture
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.29746 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 25833.xml