Intervention with selution SLR™ Agent Balloon for Endovascular Latent Limus therapy for failing AV Fistulas (ISABELLA) Trial: Protocol for a pilot clinical study and pre-clinical results. Issue 2 (March 2023)
- Record Type:
- Journal Article
- Title:
- Intervention with selution SLR™ Agent Balloon for Endovascular Latent Limus therapy for failing AV Fistulas (ISABELLA) Trial: Protocol for a pilot clinical study and pre-clinical results. Issue 2 (March 2023)
- Main Title:
- Intervention with selution SLR™ Agent Balloon for Endovascular Latent Limus therapy for failing AV Fistulas (ISABELLA) Trial: Protocol for a pilot clinical study and pre-clinical results
- Authors:
- Tang, Tjun Yip
Chong, Tze-Tec
Yap, Charyl Jia Qi
Soon, Shereen Xue Yun
Chan, Sze Ling
Tan, Ru Yu
Yap, Hao Yun
Tay, Hsien Ts'ung
Tan, Chieh-Suai
Barnhill, Scott
Hellinga, David
DeGraw, R Tyler
Finn, Aloke V - Abstract:
- Background: The aim of this pilot clinical study is to evaluate the safety and efficacy of the Selution Sustained Limus Release (SLR) ™ sirolimus-coated balloon (M.A. MedAlliance SA, Nyon, Switzerland) for improving the patency of failing arterio-venous fistulas (AVF) in hemodialysis patients. We also present herein a pre-clinical pharmacokinetic and safety evaluation of Selution™ to justify its first use in hemodialysis patients for endovascular access salvage. Methods and results: This is an investigator-initiated prospective single-center, non-blinded single-arm trial. Forty patients with clinically significant de novo or recurrent stenoses in a mature AVF circuit will be recruited. All stenotic lesions will be prepared with high pressure non-compliant conventional balloon angioplasty (CBA) prior to deployment of the Sustained-Release Selution™ sirolimus drug-eluting balloon. The primary efficacy endpoint is 6-month target lesion primary patency and the primary safety endpoint is freedom from localized or systemic serious adverse events through 30 days. Secondary endpoints of interest include technical and clinical success rates and circuit access patency at 3 and 6 months. Follow-up will occur for 2 years for those patients whose AVFs remain patent. Pharmacokinetic and histological animal safety studies performed with the Selution™ coating formulation showed prolonged arterial tissue retention of sirolimus with therapeutic levels up to 60 days and non-toxic and rapidlyBackground: The aim of this pilot clinical study is to evaluate the safety and efficacy of the Selution Sustained Limus Release (SLR) ™ sirolimus-coated balloon (M.A. MedAlliance SA, Nyon, Switzerland) for improving the patency of failing arterio-venous fistulas (AVF) in hemodialysis patients. We also present herein a pre-clinical pharmacokinetic and safety evaluation of Selution™ to justify its first use in hemodialysis patients for endovascular access salvage. Methods and results: This is an investigator-initiated prospective single-center, non-blinded single-arm trial. Forty patients with clinically significant de novo or recurrent stenoses in a mature AVF circuit will be recruited. All stenotic lesions will be prepared with high pressure non-compliant conventional balloon angioplasty (CBA) prior to deployment of the Sustained-Release Selution™ sirolimus drug-eluting balloon. The primary efficacy endpoint is 6-month target lesion primary patency and the primary safety endpoint is freedom from localized or systemic serious adverse events through 30 days. Secondary endpoints of interest include technical and clinical success rates and circuit access patency at 3 and 6 months. Follow-up will occur for 2 years for those patients whose AVFs remain patent. Pharmacokinetic and histological animal safety studies performed with the Selution™ coating formulation showed prolonged arterial tissue retention of sirolimus with therapeutic levels up to 60 days and non-toxic and rapidly declining blood levels. Histological results in animal models demonstrated safety, freedom from intraluminal thrombus, reduction in restenosis by sirolimus elution compared to CBA, and no evidence of embolic phenomena indicative of adverse particulate effects. Discussion: Long release sirolimus coated balloons may serve as a promising novel alternative therapy to paclitaxel-based technology for treating conduit stenosis secondary to neointimal hyperplasia. Pre-clinical pharmacokinetic and histological animal data are encouraging and provide suggestion of safety and efficacy in this setting. This single-center trial will provide a first step toward demonstration of efficacy and safety of this device for treatment of stenotic fistulas. … (more)
- Is Part Of:
- Journal of vascular access. Volume 24:Issue 2(2023)
- Journal:
- Journal of vascular access
- Issue:
- Volume 24:Issue 2(2023)
- Issue Display:
- Volume 24, Issue 2 (2023)
- Year:
- 2023
- Volume:
- 24
- Issue:
- 2
- Issue Sort Value:
- 2023-0024-0002-0000
- Page Start:
- 289
- Page End:
- 299
- Publication Date:
- 2023-03
- Subjects:
- Sirolimus coated balloon -- target lesion primary patency -- arterio-venous fistula -- outcome -- safety -- pre-clinical animal model
Arterial catheterization -- Periodicals
Intravenous catheterization -- Periodicals
612.13 - Journal URLs:
- http://journals.sagepub.com/home/jva ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/11297298211020867 ↗
- Languages:
- English
- ISSNs:
- 1129-7298
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25828.xml