Is faster better? A randomised crossover study comparing algorithms for closed-loop automatic oxygen control. Issue 4 (16th September 2019)
- Record Type:
- Journal Article
- Title:
- Is faster better? A randomised crossover study comparing algorithms for closed-loop automatic oxygen control. Issue 4 (16th September 2019)
- Main Title:
- Is faster better? A randomised crossover study comparing algorithms for closed-loop automatic oxygen control
- Authors:
- Schwarz, Christoph E
Kidszun, André
Bieder, Nicole S
Franz, Axel R
König, Jochem
Mildenberger, Eva
Poets, Christian F
Seyfang, Andreas
Urschitz, Michael S - Abstract:
- Abstract : Objective: Closed-loop automatic control (CLAC) of the fractional inspired oxygen (FiO2 ) improved oxygen administration to preterm infants on respiratory support. We investigated whether a revised CLAC algorithm (CLACfast, ≤2 FiO2 adjustments/min), compared with routine manual control (RMConly ), increased the proportion of time with arterial haemoglobin oxygen saturation measured by pulse oximetry within prespecified target ranges (Target%) while not being inferior to the original algorithm (CLACslow : ≤0.3 FiO2 adjustments/min). Design: Unblinded randomised controlled crossover study comparing three modes of FiO2 control in random order for 8 hours each: RMC supported by CLACfast was compared with RMConly and RMC supported by CLACslow . A computer-generated list of random numbers using a block size of six was used for the allocation sequence. Setting: Two German tertiary university neonatal intensive care units. Patients: Of 23 randomised patients, 19 were analysed (mean±SD gestational age 27±2 weeks; age at randomisation 24±10 days) on non-invasive (n=18) or invasive (n=1) respiratory support at FiO2 >0.21. Main outcome measure: Target%. Results: Mean±SD [95% CI] Target% was 68%±11% [65% to 71%] for CLACfast versus 65%±11% [61% to 68%] for CLACslow versus 58%±11% [55% to 62%] for RMConly . Prespecified hypothesis tests of: (A) superiority of CLACfast versus RMConly and (B) non-inferiority of CLACfast versus CLACslow with margin of 5% yielded one-sided p valuesAbstract : Objective: Closed-loop automatic control (CLAC) of the fractional inspired oxygen (FiO2 ) improved oxygen administration to preterm infants on respiratory support. We investigated whether a revised CLAC algorithm (CLACfast, ≤2 FiO2 adjustments/min), compared with routine manual control (RMConly ), increased the proportion of time with arterial haemoglobin oxygen saturation measured by pulse oximetry within prespecified target ranges (Target%) while not being inferior to the original algorithm (CLACslow : ≤0.3 FiO2 adjustments/min). Design: Unblinded randomised controlled crossover study comparing three modes of FiO2 control in random order for 8 hours each: RMC supported by CLACfast was compared with RMConly and RMC supported by CLACslow . A computer-generated list of random numbers using a block size of six was used for the allocation sequence. Setting: Two German tertiary university neonatal intensive care units. Patients: Of 23 randomised patients, 19 were analysed (mean±SD gestational age 27±2 weeks; age at randomisation 24±10 days) on non-invasive (n=18) or invasive (n=1) respiratory support at FiO2 >0.21. Main outcome measure: Target%. Results: Mean±SD [95% CI] Target% was 68%±11% [65% to 71%] for CLACfast versus 65%±11% [61% to 68%] for CLACslow versus 58%±11% [55% to 62%] for RMConly . Prespecified hypothesis tests of: (A) superiority of CLACfast versus RMConly and (B) non-inferiority of CLACfast versus CLACslow with margin of 5% yielded one-sided p values of <0.001 for both comparisons. Conclusions: This revised and faster CLAC algorithm was still superior to routine care in infants on respiratory support and not inferior to a previously tested slower algorithm. Trial registration number: NCT03163108 . … (more)
- Is Part Of:
- Archives of disease in childhood. Volume 105:Issue 4(2020)
- Journal:
- Archives of disease in childhood
- Issue:
- Volume 105:Issue 4(2020)
- Issue Display:
- Volume 105, Issue 4 (2020)
- Year:
- 2020
- Volume:
- 105
- Issue:
- 4
- Issue Sort Value:
- 2020-0105-0004-0000
- Page Start:
- 369
- Page End:
- 374
- Publication Date:
- 2019-09-16
- Subjects:
- Neonatology -- Respiratory -- controller -- ventilation -- hypoxia
Infants -- Diseases -- Periodicals
Newborn infants -- Diseases -- Periodicals
Fetus -- Diseases -- Periodicals
618.920105 - Journal URLs:
- http://fn.bmjjournals.com ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/archdischild-2019-317029 ↗
- Languages:
- English
- ISSNs:
- 1359-2998
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 25719.xml