An investigation into the dose rate and photon energy dependence of the GENA gel dosimeter in the MeV range. (February 2023)
- Record Type:
- Journal Article
- Title:
- An investigation into the dose rate and photon energy dependence of the GENA gel dosimeter in the MeV range. (February 2023)
- Main Title:
- An investigation into the dose rate and photon energy dependence of the GENA gel dosimeter in the MeV range
- Authors:
- Abtahi, Seyed Mohammad Mahdi
Bahrami, Farbod
Sardari, Dariush - Abstract:
- Highlights: Photon energy and dose rate dependence of GENA GD was assessed. Sensitivity of GENA GD is independent of photon energy with variations less than 2.3%. Sensitivity of GENA GD is independent of dose rate with variations less than 2.2%. Dose resolution is independent of energy and dose rate with variations less than 1.5%. Abstract: Purpose: In the current study, the energy and dose rate dependence of a new genipin-based gel dosimeter, named GENA gel dosimeter, were investigated. Methods: Prepared gel dosimeters exposed using a Varian clinical linac. Beam qualities of 6 and 18 MV were applied to investigate the GENA gel dosimeter's energy dependence. Furthermore, the gel dosimeters were exposed to 50, 100, 200, and 350 cGy/min dose rates, ranging from 0 to 8 Gy. The irradiated gel dosimeters were read out using a double beam UV–Visible spectrophotometer. The absorbance peak (AP) and area under spectrum (AUS) were evaluated. Results: Absorbance-dose sensitivities of (8.0 ± 0.18) × 10 –3 cm –1 Gy −1 and (7.8 ± 0.15) × 10 –3 cm −1 Gy −1 were obtained for GENA gel dosimeter for 6 and 18 MV beam qualities, respectively. Results specified no significant difference (p > 0.05) between the GENA gel dosimeter's sensitivities irradiated using the two energies mentioned above. For the mentioned dose rates, AP-dose sensitivities of (8.2 ± 0.22) × 10 −3, (8.1 ± 0.21) × 10 −3, (8.1 ± 0.2) × 10 −3 and (8.0 ± 0.18) × 10 −3 cm −1 Gy −1 were obtained, respectively. Results showed noHighlights: Photon energy and dose rate dependence of GENA GD was assessed. Sensitivity of GENA GD is independent of photon energy with variations less than 2.3%. Sensitivity of GENA GD is independent of dose rate with variations less than 2.2%. Dose resolution is independent of energy and dose rate with variations less than 1.5%. Abstract: Purpose: In the current study, the energy and dose rate dependence of a new genipin-based gel dosimeter, named GENA gel dosimeter, were investigated. Methods: Prepared gel dosimeters exposed using a Varian clinical linac. Beam qualities of 6 and 18 MV were applied to investigate the GENA gel dosimeter's energy dependence. Furthermore, the gel dosimeters were exposed to 50, 100, 200, and 350 cGy/min dose rates, ranging from 0 to 8 Gy. The irradiated gel dosimeters were read out using a double beam UV–Visible spectrophotometer. The absorbance peak (AP) and area under spectrum (AUS) were evaluated. Results: Absorbance-dose sensitivities of (8.0 ± 0.18) × 10 –3 cm –1 Gy −1 and (7.8 ± 0.15) × 10 –3 cm −1 Gy −1 were obtained for GENA gel dosimeter for 6 and 18 MV beam qualities, respectively. Results specified no significant difference (p > 0.05) between the GENA gel dosimeter's sensitivities irradiated using the two energies mentioned above. For the mentioned dose rates, AP-dose sensitivities of (8.2 ± 0.22) × 10 −3, (8.1 ± 0.21) × 10 −3, (8.1 ± 0.2) × 10 −3 and (8.0 ± 0.18) × 10 −3 cm −1 Gy −1 were obtained, respectively. Results showed no significant difference (p > 0.05) between the GENA gel dosimeter's sensitivities for the investigated dose rates and energies. In addition, results revealed that when the incident photon energy and dose rate changed, there were no significant differences (p > 0.05) between the GENA gel dosimeter's dose resolution values. Conclusions: It is concluded that the response of the GENA gel dosimeter is not dependent on the energy and dose rate (p > 0.05) within the studied energy and dose rate ranges. … (more)
- Is Part Of:
- Physica medica. Volume 106(2023)
- Journal:
- Physica medica
- Issue:
- Volume 106(2023)
- Issue Display:
- Volume 106, Issue 2023 (2023)
- Year:
- 2023
- Volume:
- 106
- Issue:
- 2023
- Issue Sort Value:
- 2023-0106-2023-0000
- Page Start:
- Page End:
- Publication Date:
- 2023-02
- Subjects:
- GENA gel dosimeter -- Radiochromic -- Dose rate -- Energy
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2022.102522 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25665.xml