Treatment planning with a 2.5 MV photon beam for radiation therapy. Issue 12 (27th October 2022)
- Record Type:
- Journal Article
- Title:
- Treatment planning with a 2.5 MV photon beam for radiation therapy. Issue 12 (27th October 2022)
- Main Title:
- Treatment planning with a 2.5 MV photon beam for radiation therapy
- Authors:
- Khaledi, Navid
Hayes, Chris
Belshaw, Louise
Grattan, Mark
Khan, Rao
Gräfe, James L. - Abstract:
- Abstract: Purpose: The shallow depth of maximum dose and higher dose fall‐off gradient of a 2.5 MV beam along the central axis that is available for imaging on linear accelerators is investigated for treatment of shallow tumors and sparing the organs at risk (OARs) beyond it. In addition, the 2.5 MV beam has an energy bridging the gap between kilo‐voltage (kV) and mega‐voltage (MV) beams for applications of dose enhancement with high atomic number ( Z ) nanoparticles. Methods: We have commissioned and utilized a MATLAB‐based, open‐source treatment planning software (TPS), matRad, for intensity‐modulated radiation therapy (IMRT) dose calculations. Treatment plans for prostate, liver, and head and neck (H&N), nasal cavity, two orbit cases, and glioblastoma multiforme (GBM) were performed and compared to a conventional 6 MV beam. Additional Monte Carlo calculations were also used for benchmarking the central axis dose. Results: Both beams had similar planning target volume (PTV) dose coverage for all cases. However, the 2.5 MV beam deposited 6%–19% less integral doses to the nasal cavity, orbit, and GBM cases than 6 MV photons. The mean dose to the heart in the liver plan was 10.5% lower for 2.5 MV beam. The difference between the doses to OARs of H&N for two beams was under 3%. Brain mean dose, brainstem, and optic chiasm max doses were, respectively, 7.5%–14.9%, 2.2%–8.1%, and 2.5%–19.0% lower for the 2.5 MV beam in the nasal cavity, orbit, and GBM plans. Conclusions: ThisAbstract: Purpose: The shallow depth of maximum dose and higher dose fall‐off gradient of a 2.5 MV beam along the central axis that is available for imaging on linear accelerators is investigated for treatment of shallow tumors and sparing the organs at risk (OARs) beyond it. In addition, the 2.5 MV beam has an energy bridging the gap between kilo‐voltage (kV) and mega‐voltage (MV) beams for applications of dose enhancement with high atomic number ( Z ) nanoparticles. Methods: We have commissioned and utilized a MATLAB‐based, open‐source treatment planning software (TPS), matRad, for intensity‐modulated radiation therapy (IMRT) dose calculations. Treatment plans for prostate, liver, and head and neck (H&N), nasal cavity, two orbit cases, and glioblastoma multiforme (GBM) were performed and compared to a conventional 6 MV beam. Additional Monte Carlo calculations were also used for benchmarking the central axis dose. Results: Both beams had similar planning target volume (PTV) dose coverage for all cases. However, the 2.5 MV beam deposited 6%–19% less integral doses to the nasal cavity, orbit, and GBM cases than 6 MV photons. The mean dose to the heart in the liver plan was 10.5% lower for 2.5 MV beam. The difference between the doses to OARs of H&N for two beams was under 3%. Brain mean dose, brainstem, and optic chiasm max doses were, respectively, 7.5%–14.9%, 2.2%–8.1%, and 2.5%–19.0% lower for the 2.5 MV beam in the nasal cavity, orbit, and GBM plans. Conclusions: This study demonstrates that the 2.5 MV beam can produce clinically relevant treatment plans, motivating future efforts for design of single‐energy LINACs. Such a machine will be capable of producing beams at this energy beneficial for low‐ and middle‐income countries, and investigations on dose enhancement from high‐ Z nanoparticles. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 23:Issue 12(2022)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 23:Issue 12(2022)
- Issue Display:
- Volume 23, Issue 12 (2022)
- Year:
- 2022
- Volume:
- 23
- Issue:
- 12
- Issue Sort Value:
- 2022-0023-0012-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2022-10-27
- Subjects:
- 2.5 MV photons -- commissioning -- IMRT -- Monte Carlo simulation -- radiotherapy -- treatment planning
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13811 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25605.xml