Validation of pencil beam scanning proton therapy with multi‐leaf collimator calculated by a commercial Monte Carlo dose engine. Issue 12 (24th November 2022)
- Record Type:
- Journal Article
- Title:
- Validation of pencil beam scanning proton therapy with multi‐leaf collimator calculated by a commercial Monte Carlo dose engine. Issue 12 (24th November 2022)
- Main Title:
- Validation of pencil beam scanning proton therapy with multi‐leaf collimator calculated by a commercial Monte Carlo dose engine
- Authors:
- Tominaga, Yuki
Sakurai, Yusuke
Miyata, Junya
Harada, Shuichi
Akagi, Takashi
Oita, Masataka - Abstract:
- Abstract: This study aimed to evaluate the clinical beam commissioning results and lateral penumbra characteristics of our new pencil beam scanning (PBS) proton therapy using a multi‐leaf collimator (MLC) calculated by use of a commercial Monte Carlo dose engine. Eighteen collimated uniform dose plans for cubic targets were optimized by the RayStation 9A treatment planning system (TPS), varying scan area, modulation widths, measurement depths, and collimator angles. To test the patient‐specific measurements, we also created and verified five clinically realistic PBS plans with the MLC, such as the liver, prostate, base‐of‐skull, C‐shape, and head‐and‐neck. The verification measurements consist of the depth dose (DD), lateral profile (LP), and absolute dose (AD). We compared the LPs and ADs between the calculation and measurements. For the cubic plans, the gamma index pass rates (γ‐passing) were on average 96.5% ± 4.0% at 3%/3 mm for the DD and 95.2% ± 7.6% at 2%/2 mm for the LP. In several LP measurements less than 75 mm depths, the γ‐passing deteriorated (increased the measured doses) by less than 90% with the scattering such as the MLC edge and range shifter. The deteriorated γ‐passing was satisfied by more than 90% at 2%/2 mm using uncollimated beams instead of collimated beams except for three planes. The AD differences and the lateral penumbra width (80%–20% distance) were within ±1.9% and ± 1.1 mm, respectively. For the clinical plan measurements, the γ‐passing of LPAbstract: This study aimed to evaluate the clinical beam commissioning results and lateral penumbra characteristics of our new pencil beam scanning (PBS) proton therapy using a multi‐leaf collimator (MLC) calculated by use of a commercial Monte Carlo dose engine. Eighteen collimated uniform dose plans for cubic targets were optimized by the RayStation 9A treatment planning system (TPS), varying scan area, modulation widths, measurement depths, and collimator angles. To test the patient‐specific measurements, we also created and verified five clinically realistic PBS plans with the MLC, such as the liver, prostate, base‐of‐skull, C‐shape, and head‐and‐neck. The verification measurements consist of the depth dose (DD), lateral profile (LP), and absolute dose (AD). We compared the LPs and ADs between the calculation and measurements. For the cubic plans, the gamma index pass rates (γ‐passing) were on average 96.5% ± 4.0% at 3%/3 mm for the DD and 95.2% ± 7.6% at 2%/2 mm for the LP. In several LP measurements less than 75 mm depths, the γ‐passing deteriorated (increased the measured doses) by less than 90% with the scattering such as the MLC edge and range shifter. The deteriorated γ‐passing was satisfied by more than 90% at 2%/2 mm using uncollimated beams instead of collimated beams except for three planes. The AD differences and the lateral penumbra width (80%–20% distance) were within ±1.9% and ± 1.1 mm, respectively. For the clinical plan measurements, the γ‐passing of LP at 2%/2 mm and the AD differences were 97.7% ± 4.2% on average and within ±1.8%, respectively. The measurements were in good agreement with the calculations of both the cubic and clinical plans inserted in the MLC except for LPs less than 75 mm regions of some cubic and clinical plans. The calculation errors in collimated beams can be mitigated by substituting uncollimated beams. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 23:Issue 12(2022)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 23:Issue 12(2022)
- Issue Display:
- Volume 23, Issue 12 (2022)
- Year:
- 2022
- Volume:
- 23
- Issue:
- 12
- Issue Sort Value:
- 2022-0023-0012-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2022-11-24
- Subjects:
- commissioning -- lateral penumbra -- multi‐leaf collimator -- pencil beam scanning -- proton therapy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13817 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25605.xml