Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist device. Issue 4 (August 2021)
- Record Type:
- Journal Article
- Title:
- Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist device. Issue 4 (August 2021)
- Main Title:
- Blood pressure from the optical Aktiia Bracelet
- Authors:
- Vybornova, Anna
Polychronopoulou, Erietta
Wurzner-Ghajarzadeh, Arlène
Fallet, Sibylle
Sola, Josep
Wuerzner, Gregoire - Abstract:
- Abstract : Objective: The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month. Methods: In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject. Results: Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP. Conclusion: After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 inAbstract : Objective: The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month. Methods: In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject. Results: Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP. Conclusion: After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population. … (more)
- Is Part Of:
- Blood pressure monitoring. Volume 26:Issue 4(2021)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 26:Issue 4(2021)
- Issue Display:
- Volume 26, Issue 4 (2021)
- Year:
- 2021
- Volume:
- 26
- Issue:
- 4
- Issue Sort Value:
- 2021-0026-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-08
- Subjects:
- blood pressure -- calibration -- clinical investigation -- cuffless blood pressure monitor -- optical signals -- initialization -- validation
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000531 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 25578.xml