A Comparative Study Between the Stryker EZout Powered Acetabular Revision System and the Zimmer Explant Acetabular Cup Removal Systems. Issue 2 (June 2021)
- Record Type:
- Journal Article
- Title:
- A Comparative Study Between the Stryker EZout Powered Acetabular Revision System and the Zimmer Explant Acetabular Cup Removal Systems. Issue 2 (June 2021)
- Main Title:
- A Comparative Study Between the Stryker EZout Powered Acetabular Revision System and the Zimmer Explant Acetabular Cup Removal Systems
- Authors:
- Kwong, Louis
Billi, Fabrizio
Keller, Scott
Kavanaugh, Aaron
Luu, Andrew
Paprosky, Wayne - Abstract:
- Abstract : The conservation of bone stock and the decrease of intraoperative acetabular fracture risk associated with revision surgery is a primary goal to improve the outcome of the procedure. The objective of this study was to evaluate and compare the performance of the Explant Acetabular Cup Removal System and the EZ out Powered Acetabular Revision System in an in vitro model. Acetabular components were implanted into instrumented composite hemipelvises and divided into 2 groups, EZ out System and Explant System. One experienced orthopedic surgeon and 1 orthopedic resident physician performed the removal procedures. The strains at various points in the periacetabular bone, the temperature at the implant-bone interface, the total time to removal, the torque applied to the implant and the amount of acetabular foam on each cup after extraction, as a surrogate for cavitary acetabular bone stock loss, were calculated for each test. Statistical analysis was conducted using 2-way multivariate analysis of variance followed by Tukey honestly significant difference multiple comparisons. The EZ out System required an overall lower force and torque ( P <0.0001, both) during removal, producing lower strains in the surrounding composite bone. The procedure was faster ( P <0.0001) and less energy demanding with the EZ out compared with the Explant. The amount of foam on the cup was on average less for the EZ out than for the Explant ( P <0.05). We found that the EZ out System isAbstract : The conservation of bone stock and the decrease of intraoperative acetabular fracture risk associated with revision surgery is a primary goal to improve the outcome of the procedure. The objective of this study was to evaluate and compare the performance of the Explant Acetabular Cup Removal System and the EZ out Powered Acetabular Revision System in an in vitro model. Acetabular components were implanted into instrumented composite hemipelvises and divided into 2 groups, EZ out System and Explant System. One experienced orthopedic surgeon and 1 orthopedic resident physician performed the removal procedures. The strains at various points in the periacetabular bone, the temperature at the implant-bone interface, the total time to removal, the torque applied to the implant and the amount of acetabular foam on each cup after extraction, as a surrogate for cavitary acetabular bone stock loss, were calculated for each test. Statistical analysis was conducted using 2-way multivariate analysis of variance followed by Tukey honestly significant difference multiple comparisons. The EZ out System required an overall lower force and torque ( P <0.0001, both) during removal, producing lower strains in the surrounding composite bone. The procedure was faster ( P <0.0001) and less energy demanding with the EZ out compared with the Explant. The amount of foam on the cup was on average less for the EZ out than for the Explant ( P <0.05). We found that the EZ out System is effective in achieving safe removal of a well-fixed acetabular component in an in vitro model of cementless fixation. This system should be considered as a reasonable alternative to manual removal techniques. … (more)
- Is Part Of:
- Techniques in orthopaedics. Volume 36:Issue 2(2021)
- Journal:
- Techniques in orthopaedics
- Issue:
- Volume 36:Issue 2(2021)
- Issue Display:
- Volume 36, Issue 2 (2021)
- Year:
- 2021
- Volume:
- 36
- Issue:
- 2
- Issue Sort Value:
- 2021-0036-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-06
- Subjects:
- hip -- revision -- revision arthroplasty -- arthroplasty -- EZout -- Explant -- acetabular cup -- bone conservation -- osteotome
Orthopedics -- Periodicals
617.47 - Journal URLs:
- http://journals.lww.com/techortho/pages/default.aspx ↗
http://www.lww.com/Product/0885-9698 ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/BTO.0000000000000409 ↗
- Languages:
- English
- ISSNs:
- 0885-9698
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8745.278000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25565.xml