A case report of inappropriate inhibition of ventricular pacing due to a unique pacemaker electrogram storage feature. Issue 3 (20th March 2021)
- Record Type:
- Journal Article
- Title:
- A case report of inappropriate inhibition of ventricular pacing due to a unique pacemaker electrogram storage feature. Issue 3 (20th March 2021)
- Main Title:
- A case report of inappropriate inhibition of ventricular pacing due to a unique pacemaker electrogram storage feature
- Authors:
- Khan, Habib R
Chan, William K
Kanawati, Juliana
Yee, Raymond - Editors:
- Sinning, Christoph
Sohns, Christian
Iliodromitis, Konstantinos
Barbieri, Fabian
Duncker, David
Ang, Richard
McNaughton, Edwina
Ahmed, Nida - Abstract:
- Abstract: Background: Modern permanent pacemakers (PPMs) have individual features designed to identify cardiac rhythm abnormalities and improve their performance. Inappropriate pacing inhibition may be an undesired outcome from these features and cause symptoms in patients who require frequent pacing, leading to dizziness, and syncope. Inappropriate inhibition can be difficult to identify in circumstances that are intermittent and difficult to reproduce. Case summary: A 57-year-old female underwent a mitral valve replacement (MVR) for severe mitral stenosis. One month following MVR, she presented with symptomatic third-degree atrioventricular block, and a dual-chamber PPM (Advisa™, Medtronic, Minneapolis, USA) was implanted and programmed DDD 50–130 b.p.m. At the 3-month follow-up, she reported frequent episodes of lightheadedness. She was found to have intermittent ventricular pacing inhibition on a 48-h Holter monitor due to an internal function of the Advisa™ series of PPMs that attempts to store an electrogram (EGM) every 1 h and 30 s. During the EGM storage, an amplified signal from the storage capacitor can result in oversensing by the ventricular channel and inappropriate pacing inhibition. Discussion: To rectify the issue, the ventricular lead sensitivity value was increased from 0.9 mV to 1.2 mV. No instances of inappropriate ventricular pacing inhibition were noted on follow-up. To our knowledge, this is a rare case of inappropriate ventricular pacing inhibitionAbstract: Background: Modern permanent pacemakers (PPMs) have individual features designed to identify cardiac rhythm abnormalities and improve their performance. Inappropriate pacing inhibition may be an undesired outcome from these features and cause symptoms in patients who require frequent pacing, leading to dizziness, and syncope. Inappropriate inhibition can be difficult to identify in circumstances that are intermittent and difficult to reproduce. Case summary: A 57-year-old female underwent a mitral valve replacement (MVR) for severe mitral stenosis. One month following MVR, she presented with symptomatic third-degree atrioventricular block, and a dual-chamber PPM (Advisa™, Medtronic, Minneapolis, USA) was implanted and programmed DDD 50–130 b.p.m. At the 3-month follow-up, she reported frequent episodes of lightheadedness. She was found to have intermittent ventricular pacing inhibition on a 48-h Holter monitor due to an internal function of the Advisa™ series of PPMs that attempts to store an electrogram (EGM) every 1 h and 30 s. During the EGM storage, an amplified signal from the storage capacitor can result in oversensing by the ventricular channel and inappropriate pacing inhibition. Discussion: To rectify the issue, the ventricular lead sensitivity value was increased from 0.9 mV to 1.2 mV. No instances of inappropriate ventricular pacing inhibition were noted on follow-up. To our knowledge, this is a rare case of inappropriate ventricular pacing inhibition caused by a combination of PPM self-adjusting sensitivity algorithm and oversensing every 1 h and 30 s from an amplified storage capacitor. Physicians should be aware of this possible complication and differentiate it from device or lead malfunction. … (more)
- Is Part Of:
- European heart journal. Volume 5:Issue 3(2021)
- Journal:
- European heart journal
- Issue:
- Volume 5:Issue 3(2021)
- Issue Display:
- Volume 5, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 5
- Issue:
- 3
- Issue Sort Value:
- 2021-0005-0003-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-03-20
- Subjects:
- Pacemaker -- Pacing inhibition -- Advisa -- Case report -- Oversensing -- Dizziness
Cardiology -- Periodicals
Cardiology -- Case studies -- Periodicals
Heart -- Diseases -- Periodicals
Heart -- Diseases -- Case studies -- Periodicals
616.12 - Journal URLs:
- http://www.oxfordjournals.org/ ↗
https://academic.oup.com/ehjcr/ ↗ - DOI:
- 10.1093/ehjcr/ytab096 ↗
- Languages:
- English
- ISSNs:
- 2514-2119
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25542.xml