Investigating the impact of collimator size variation on the single beam radiation of Gamma Knife PerfexionTM based on Monte Carlo simulation. (March 2020)
- Record Type:
- Journal Article
- Title:
- Investigating the impact of collimator size variation on the single beam radiation of Gamma Knife PerfexionTM based on Monte Carlo simulation. (March 2020)
- Main Title:
- Investigating the impact of collimator size variation on the single beam radiation of Gamma Knife PerfexionTM based on Monte Carlo simulation
- Authors:
- Junios,
Irhas,
Novitrian,
Soediatmoko, E
Haryanto, F
Su'ud, Z
Fielding, A.L. - Abstract:
- Abstract: The Gamma Knife Perfexion TM (Elekta Oncology Systems) is a specialized stereotactic radiotherapy device that uses 192 Cobalt-60 radioactive sources. A dynamic collimator system enables three circular field sizes of 4 mm, 8 mm, and 16 mm diameter to be delivered. This study investigates the dosimetric characteristics of the three different sized collimated beams using Monte Carlo (MC) simulations. The MC simulation made use of previously calculated phase space from a Monte-Carlo model of the encapsulated Cobalt-60 sources. The collimators were modelled using the FLATFILT component module of the BEAMnrc user code. The output of the single beam radiation was stored in a circular scoring plane with a radius of 1.4 cm at source surface distance (SSD) 17.20 cm in the phase space file. The simulation parameters were used 2.1 x 10 9 particles, electron cut off energy (ECUT) of 0.7 MeV, and photon cut off energy (PCUT) of 0.01 MeV. To investigate the physical full width at half maximum (FWHM) and penumbra width (80-20%), the single beam profiles and mean energy were obtained using BEAMnrc/BEAMDP code. Based on the results that there was an increase of the full width at half maximum (FWHM) with the size of the collimator on the beam profile and the mean energy. The FHWM of the energy fluence profile for the collimator 4, 8, and 16 mm are 2.22, 3.90, and 7.80 mm, respectively. The FHWM of the fluence profile for the collimator sizes 4, 8, and 16 mm are 1.87, 3.87, and 7.80Abstract: The Gamma Knife Perfexion TM (Elekta Oncology Systems) is a specialized stereotactic radiotherapy device that uses 192 Cobalt-60 radioactive sources. A dynamic collimator system enables three circular field sizes of 4 mm, 8 mm, and 16 mm diameter to be delivered. This study investigates the dosimetric characteristics of the three different sized collimated beams using Monte Carlo (MC) simulations. The MC simulation made use of previously calculated phase space from a Monte-Carlo model of the encapsulated Cobalt-60 sources. The collimators were modelled using the FLATFILT component module of the BEAMnrc user code. The output of the single beam radiation was stored in a circular scoring plane with a radius of 1.4 cm at source surface distance (SSD) 17.20 cm in the phase space file. The simulation parameters were used 2.1 x 10 9 particles, electron cut off energy (ECUT) of 0.7 MeV, and photon cut off energy (PCUT) of 0.01 MeV. To investigate the physical full width at half maximum (FWHM) and penumbra width (80-20%), the single beam profiles and mean energy were obtained using BEAMnrc/BEAMDP code. Based on the results that there was an increase of the full width at half maximum (FWHM) with the size of the collimator on the beam profile and the mean energy. The FHWM of the energy fluence profile for the collimator 4, 8, and 16 mm are 2.22, 3.90, and 7.80 mm, respectively. The FHWM of the fluence profile for the collimator sizes 4, 8, and 16 mm are 1.87, 3.87, and 7.80 mm. The FHWM of mean energy as a function of position for the collimator sizes 4, 8, and 16 mm are 2.56, 4.20, and 7.94 mm, respectively. The maximum mean energy for the collimators decreases as much as 0.0185 MeV. … (more)
- Is Part Of:
- Journal of physics. Volume 1505(2020)
- Journal:
- Journal of physics
- Issue:
- Volume 1505(2020)
- Issue Display:
- Volume 1505, Issue 1 (2020)
- Year:
- 2020
- Volume:
- 1505
- Issue:
- 1
- Issue Sort Value:
- 2020-1505-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-03
- Subjects:
- Physics -- Congresses
530.5 - Journal URLs:
- http://www.iop.org/EJ/journal/1742-6596 ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1742-6596/1505/1/012013 ↗
- Languages:
- English
- ISSNs:
- 1742-6588
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5036.223000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 25446.xml