Smartphone application-supported validation of three automatic devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010: the Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B. Issue 6 (December 2021)
- Record Type:
- Journal Article
- Title:
- Smartphone application-supported validation of three automatic devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010: the Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B. Issue 6 (December 2021)
- Main Title:
- Smartphone application-supported validation of three automatic devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010
- Authors:
- Zhang, Puhong
Li, Xi
Fang, Zhe
Lu, Yanling
Cui, Jingchen
Du, Xin
Hu, Rong - Abstract:
- Abstract : Background: Accurate measurement of blood pressure (BP) is crucial to hypertension control and prevention of future stroke and heart attack. All BP measuring devices must be validated independently in the clinical setting. Objective: To validate the accuracy of three automatic upper arm devices (Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B) for self-measurement of BP in Chinese adults with arm size of 22–32 cm. Methods: The validation was conducted independently for each of the three devices according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP revision 2010), with the facilitation of a designated smartphone application. Subjects were recruited from those attending Beijing Anzhen Hospital for routine physical examination and clinic visits. For each device, BP was measured sequentially in 33 adults using a mercury sphygmomanometer (two observers) and the test device (one supervisor) with seven measurements alternating between observers and the device, which generated a total of 99 before/after paired values for SBP and DBP separately. The judgments were made based on the distribution of the paired difference among the 99 measurements (Part 1) and among the 33 subjects (Part 2). To pass, a device must achieve all the minimum Pass requirements in Part 1 and Part 2 for both SBP and DBP (Part 3). Results: Only HEM-7120 achieved the part 1 and part 2 targets for both SBP and DBP. KF-65B achieved the DBP targets of part 1 andAbstract : Background: Accurate measurement of blood pressure (BP) is crucial to hypertension control and prevention of future stroke and heart attack. All BP measuring devices must be validated independently in the clinical setting. Objective: To validate the accuracy of three automatic upper arm devices (Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B) for self-measurement of BP in Chinese adults with arm size of 22–32 cm. Methods: The validation was conducted independently for each of the three devices according to the European Society of Hypertension International Protocol revision 2010 (ESH-IP revision 2010), with the facilitation of a designated smartphone application. Subjects were recruited from those attending Beijing Anzhen Hospital for routine physical examination and clinic visits. For each device, BP was measured sequentially in 33 adults using a mercury sphygmomanometer (two observers) and the test device (one supervisor) with seven measurements alternating between observers and the device, which generated a total of 99 before/after paired values for SBP and DBP separately. The judgments were made based on the distribution of the paired difference among the 99 measurements (Part 1) and among the 33 subjects (Part 2). To pass, a device must achieve all the minimum Pass requirements in Part 1 and Part 2 for both SBP and DBP (Part 3). Results: Only HEM-7120 achieved the part 1 and part 2 targets for both SBP and DBP. KF-65B achieved the DBP targets of part 1 and part 2 but failed for SBP. YE680A only achieved the DBP targets of part 2 but failed for all others. The findings also indicated that the devices had higher SBP readings (1.3 mmHg, 1.0 mmHg and 4.1 mmHg higher for HEM-7120, YE680A and KF-65B, respectively) and lower DBP readings (2.0 mmHg, 1.1 mmHg and 3.3 mmHg lower, respectively) when compared to the mercury sphygmomanometer. Conclusions: The Omron HEM-7120 passed the requirements of the ESH-IP 2010 revision, while the Yuwell YE680A and Cofoe KF-65B failed (part 3). … (more)
- Is Part Of:
- Blood pressure monitoring. Volume 26:Issue 6(2021)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 26:Issue 6(2021)
- Issue Display:
- Volume 26, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 26
- Issue:
- 6
- Issue Sort Value:
- 2021-0026-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-12
- Subjects:
- blood pressure measurement -- device -- European Society of Hypertension International Protocol revision 2010 -- upper arm
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000547 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 25338.xml