Comparison of platelet count results on the Sysmex XN between citrate or MgSO4 and K2 EDTA anticoagulants. (12th September 2022)
- Record Type:
- Journal Article
- Title:
- Comparison of platelet count results on the Sysmex XN between citrate or MgSO4 and K2 EDTA anticoagulants. (12th September 2022)
- Main Title:
- Comparison of platelet count results on the Sysmex XN between citrate or MgSO4 and K2 EDTA anticoagulants
- Authors:
- Soulard, Michel
Croix, Pascale
Cohen, Patrick - Abstract:
- Abstract: Introduction: The aim of this study performed on Sysmex XN is to compare platelet values on citrate and MgSO4 (TBX) in patients with K2EDTA‐induced platelet clusters and to identify platelet biases of these matrices compared to K2EDTA. Methods: Sixty patients with K2EDTA‐induced platelet clusters were re‐sampled with K2EDTA, citrate and TBX. Platelet results were then compared, and smears were analysed for clumping. Platelet results from 120 patients without K2EDTA‐induced platelet clusters were compared between K2 EDTA, citrate, and MgSO4 with impedance and fluorescence modes. Biases from regressions were analysed. Results: Out of the 60 patients with K2EDTA‐induced platelet clusters, none showed platelet clusters with MgSO4 whereas 50% still showed clusters with citrate. Among those without platelet clusters on citrate, the mean relative difference between (citrate‐ MgSO4 )/MgSO4 was −12.7% in impedance and −9.8% in fluorescence. Among the 120 patients without K2EDTA‐induced platelet clusters, in fluorescence the mean relative bias with respect to K2EDTA was −2.06% for MgSO4 and −10.3% for Citrate. For the MgSO4 versus K2 EDTA regressions, the maximum absolute values of the 95% CI of the relative biases at 150 × 10 9 /L (5.45%) and 450 × 10 9 /L (3.56%) were below the desirable analytical objectives of the EFLM. Conclusion: In patients with K2EDTA‐induced platelet clusters, MgSO4 is preferable to citrate. MgSO4 provides a bias with XN in fluorescence whenAbstract: Introduction: The aim of this study performed on Sysmex XN is to compare platelet values on citrate and MgSO4 (TBX) in patients with K2EDTA‐induced platelet clusters and to identify platelet biases of these matrices compared to K2EDTA. Methods: Sixty patients with K2EDTA‐induced platelet clusters were re‐sampled with K2EDTA, citrate and TBX. Platelet results were then compared, and smears were analysed for clumping. Platelet results from 120 patients without K2EDTA‐induced platelet clusters were compared between K2 EDTA, citrate, and MgSO4 with impedance and fluorescence modes. Biases from regressions were analysed. Results: Out of the 60 patients with K2EDTA‐induced platelet clusters, none showed platelet clusters with MgSO4 whereas 50% still showed clusters with citrate. Among those without platelet clusters on citrate, the mean relative difference between (citrate‐ MgSO4 )/MgSO4 was −12.7% in impedance and −9.8% in fluorescence. Among the 120 patients without K2EDTA‐induced platelet clusters, in fluorescence the mean relative bias with respect to K2EDTA was −2.06% for MgSO4 and −10.3% for Citrate. For the MgSO4 versus K2 EDTA regressions, the maximum absolute values of the 95% CI of the relative biases at 150 × 10 9 /L (5.45%) and 450 × 10 9 /L (3.56%) were below the desirable analytical objectives of the EFLM. Conclusion: In patients with K2EDTA‐induced platelet clusters, MgSO4 is preferable to citrate. MgSO4 provides a bias with XN in fluorescence when compared to EDTA which is within analytical tolerance. … (more)
- Is Part Of:
- International journal of laboratory hematology. Volume 45:Number 1(2023)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 45:Number 1(2023)
- Issue Display:
- Volume 45, Issue 1 (2023)
- Year:
- 2023
- Volume:
- 45
- Issue:
- 1
- Issue Sort Value:
- 2023-0045-0001-0000
- Page Start:
- 20
- Page End:
- 28
- Publication Date:
- 2022-09-12
- Subjects:
- citrate -- ethylene diamine tetracetic acid -- magnesium sulphate -- platelet -- Sysmex
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.13966 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 25050.xml