Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices. (20th September 2021)
- Record Type:
- Journal Article
- Title:
- Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices. (20th September 2021)
- Main Title:
- Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices
- Authors:
- Matsubara, Hiroaki
Ezura, Takatomo
Hashimoto, Yaichiro
Karasawa, Kumiko
Nishio, Teiji
Tsuneda, Masato - Abstract:
- Abstract: Cardiac implantable electronic devices (CIEDs) were believed to have a tolerance dose and that direct irradiation has to be avoided. Thus, no clinical guidelines have mentioned the feasibility of total body irradiation (TBI) with a CIED directly. The purpose of this work was to study a feasible and safe condition for TBI using a CIED. Eighteen CIEDs were directly irradiated by a 6-MV X-ray beam, where a non-neutron producible beam was employed for the removal of any neutron contribution to CIED malfunction. Irradiation up to 10 Gy in accumulated dose was conducted with a 100-cGy/min dose rate, followed by up to 20 Gy at 200 cGy/min. An irradiation test of whether inappropriate ventricular shock therapy was triggered or not was also performed by using a 6-MV beam of 5, 10, 20 and 40 cGy/min to two CIEDs. No malfunction was observed during irradiation up to 20 Gy at 100 and 200 cGy/min without activation of shock therapy. These results were compared with typical TBI, suggesting that a CIED in TBI will not encounter malfunction because the prescribed dose and the dose rate required for TBI are much safer than those used in this experiment. Several inappropriate shock therapies were, however, observed even at 10 cGy/min if activated. The present result suggested that TBI was feasible and safe if a non-neutron producible beam was employed at low dose-rate without activation of shock therapy, where it was not inconsistent with clinical and non-clinical data in theAbstract: Cardiac implantable electronic devices (CIEDs) were believed to have a tolerance dose and that direct irradiation has to be avoided. Thus, no clinical guidelines have mentioned the feasibility of total body irradiation (TBI) with a CIED directly. The purpose of this work was to study a feasible and safe condition for TBI using a CIED. Eighteen CIEDs were directly irradiated by a 6-MV X-ray beam, where a non-neutron producible beam was employed for the removal of any neutron contribution to CIED malfunction. Irradiation up to 10 Gy in accumulated dose was conducted with a 100-cGy/min dose rate, followed by up to 20 Gy at 200 cGy/min. An irradiation test of whether inappropriate ventricular shock therapy was triggered or not was also performed by using a 6-MV beam of 5, 10, 20 and 40 cGy/min to two CIEDs. No malfunction was observed during irradiation up to 20 Gy at 100 and 200 cGy/min without activation of shock therapy. These results were compared with typical TBI, suggesting that a CIED in TBI will not encounter malfunction because the prescribed dose and the dose rate required for TBI are much safer than those used in this experiment. Several inappropriate shock therapies were, however, observed even at 10 cGy/min if activated. The present result suggested that TBI was feasible and safe if a non-neutron producible beam was employed at low dose-rate without activation of shock therapy, where it was not inconsistent with clinical and non-clinical data in the literature. The feasibility of TBI while using a CIED was discussed for the first time. … (more)
- Is Part Of:
- Journal of radiation research. Volume 62:Number 6(2021)
- Journal:
- Journal of radiation research
- Issue:
- Volume 62:Number 6(2021)
- Issue Display:
- Volume 62, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 62
- Issue:
- 6
- Issue Sort Value:
- 2021-0062-0006-0000
- Page Start:
- 1006
- Page End:
- 1014
- Publication Date:
- 2021-09-20
- Subjects:
- cardiac implantable electronic devices (CIED) -- pacemaker (PM) -- implantable cardioverter defibrillator (ICD) -- cardiac resynchronization therapy (CRT) -- total body irradiation (TBI)
Radiology, Medical -- Periodicals
Radiobiology -- Periodicals
Radiation -- Periodicals
616.0757 - Journal URLs:
- http://bibpurl.oclc.org/web/15847 ↗
http://bibpurl.oclc.org/web/7828 ↗
http://www.journalarchive.jst.go.jp/english/jnltop_en.php?cdjournal=jrr1960 ↗
https://www.jstage.jst.go.jp/browse/jrr ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/jrr/rrab088 ↗
- Languages:
- English
- ISSNs:
- 0449-3060
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24976.xml