Differences in lung clearance index and functional residual capacity between two commercial multiple-breath nitrogen washout devices in healthy children and adults. Issue 2 (29th June 2020)
- Record Type:
- Journal Article
- Title:
- Differences in lung clearance index and functional residual capacity between two commercial multiple-breath nitrogen washout devices in healthy children and adults. Issue 2 (29th June 2020)
- Main Title:
- Differences in lung clearance index and functional residual capacity between two commercial multiple-breath nitrogen washout devices in healthy children and adults
- Authors:
- Zwitserloot, Annelies M.
van den Born, Evelyne J.
Raaijmakers, Lena H.A.
Stalman, Wouter E.
van Smaalen, Marjanne
van den Berge, Maarten
Gappa, Monika
Koppelman, Gerard H.
Willemse, Brigitte W.M. - Abstract:
- Multiple-breath nitrogen washout (MBNW) and its clinical parameter lung clearance index (LCI) are gaining increasing attention for the assessment of small airway function. Measurement of LCI relies on accurate assessment of functional residual capacity (FRC). The EasyOne Pro LAB (ndd) and Exhalyzer D (EM) are two commercially available MBNW devices. The aim of the study was to compare these two devices in vitro and in vivo in healthy subjects with regard to FRC, LCI and secondary outcome parameters and to relate FRCMBNW to FRC measured by body plethysmography (pleth) and helium dilution technique. MBNW measurements were performed using a lung model (FRC between 500 and 4000 mL) in vitro and in 38 subjects aged 6–65 years followed by helium dilution and pleth in vivo using fixed and relaxed breathing techniques. In vitro accuracy within 5% of lung model FRC was 67.3% for ndd, FRC was >5% higher for EM in all tests. In vivo, FRCpleth ranged from 1.2 to 5.6 L. Mean differences (limits of agreement) between FRCpleth and FRCMBNW were −7.0%, (−23.2 to 9.2%) and 5.7% (−11.2 to 22.6%) using ndd and EM, respectively. FRCndd was consistently lower than FRCEM (−11.8% (−25.6 to 2%)). LCI was comparable between the two devices (−1.3% (−21.9 to 19.3%)). There was a difference of >10 % in LCI in 12 of 38 subjects. Using the most recent software updates, both devices show relevant deviations in FRC measurement both in vitro and in vivo and individual differences in LCI in a significantMultiple-breath nitrogen washout (MBNW) and its clinical parameter lung clearance index (LCI) are gaining increasing attention for the assessment of small airway function. Measurement of LCI relies on accurate assessment of functional residual capacity (FRC). The EasyOne Pro LAB (ndd) and Exhalyzer D (EM) are two commercially available MBNW devices. The aim of the study was to compare these two devices in vitro and in vivo in healthy subjects with regard to FRC, LCI and secondary outcome parameters and to relate FRCMBNW to FRC measured by body plethysmography (pleth) and helium dilution technique. MBNW measurements were performed using a lung model (FRC between 500 and 4000 mL) in vitro and in 38 subjects aged 6–65 years followed by helium dilution and pleth in vivo using fixed and relaxed breathing techniques. In vitro accuracy within 5% of lung model FRC was 67.3% for ndd, FRC was >5% higher for EM in all tests. In vivo, FRCpleth ranged from 1.2 to 5.6 L. Mean differences (limits of agreement) between FRCpleth and FRCMBNW were −7.0%, (−23.2 to 9.2%) and 5.7% (−11.2 to 22.6%) using ndd and EM, respectively. FRCndd was consistently lower than FRCEM (−11.8% (−25.6 to 2%)). LCI was comparable between the two devices (−1.3% (−21.9 to 19.3%)). There was a difference of >10 % in LCI in 12 of 38 subjects. Using the most recent software updates, both devices show relevant deviations in FRC measurement both in vitro and in vivo and individual differences in LCI in a significant proportion of subjects. The devices are therefore not interchangeable. MBNW measurements with the Exhalyzer D and EasyOne Pro LAB cannot be used interchangeably for FRC and LCI measurements. FRC measured on both devices showed deviations from in vitro and in vivo measurements. https://bit.ly/2xyyUuJ … (more)
- Is Part Of:
- ERJ open research. Volume 6:Issue 2(2020)
- Journal:
- ERJ open research
- Issue:
- Volume 6:Issue 2(2020)
- Issue Display:
- Volume 6, Issue 2 (2020)
- Year:
- 2020
- Volume:
- 6
- Issue:
- 2
- Issue Sort Value:
- 2020-0006-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-06-29
- Subjects:
- Respiratory organs -- Diseases -- Periodicals
Respiration -- Periodicals
Respiration
Respiratory organs -- Diseases
Respiratory organs -- Diseases -- Treatment
Respiratory Tract Diseases
Electronic journals
Fulltext
Internet Resources
Periodicals
Periodical
616.2005 - Journal URLs:
- http://openres.ersjournals.com/ ↗
http://bibpurl.oclc.org/web/76947 ↗ - DOI:
- 10.1183/23120541.00247-2019 ↗
- Languages:
- English
- ISSNs:
- 2312-0541
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 24854.xml