Cardiac tamponade as an inherent but potentially nonfatal complication of transvenous lead extraction: Experience with 1126 procedures performed using mechanical tools. (14th September 2022)
- Record Type:
- Journal Article
- Title:
- Cardiac tamponade as an inherent but potentially nonfatal complication of transvenous lead extraction: Experience with 1126 procedures performed using mechanical tools. (14th September 2022)
- Main Title:
- Cardiac tamponade as an inherent but potentially nonfatal complication of transvenous lead extraction: Experience with 1126 procedures performed using mechanical tools
- Authors:
- Nowosielecka, Dorota
Tułecki, Łukasz
Jacheć, Wojciech
Polewczyk, Anna
Tomków, Konrad
Stefańczyk, Paweł
Bródka, Jarosław
Kutarski, Andrzej - Abstract:
- Abstract: Introduction: Cardiac tamponade (CT) is one of the most common and dangerous complications of transvenous lead extraction (TLE). So far, however, there has been little discussion about the problem. Methods: We analyzed the occurrence of CT in a group of 1226 patients undergoing TLE at a single reference center between June, 2015 and February, 2021. Using standard mechanical devices as first‐line tools, a total of 2092 leads had been extracted. Results: CT occurred in 18 patients (1.47%): due to injury to the wall of the right atrium in 14 patients (1.14%) and other cardiac walls in four patients (0.33%). Younger patient age at first implantation, female gender, high left ventricular ejection fraction (LVEF), lower New York Heart Association class, low Charlson comorbidity index, longer implant duration, and the number of previous procedures related to cardiac implantable electronic devices (CIED) are important patient‐related risk factors for CT. Significant procedure‐related risk factors include the number of extracted leads, extraction of atrial leads and longer dwell time of extracted leads. Intraoperative transoesophageal echocardiography (TEE) provides a lot of information about pulling on various cardiac structures and is able to detect a very early phase of bleeding to the pericardial sac. As a result of implementing best practices guidance in performing extraction procedures and close collaboration with cardiac surgeons that allowed immediate rescueAbstract: Introduction: Cardiac tamponade (CT) is one of the most common and dangerous complications of transvenous lead extraction (TLE). So far, however, there has been little discussion about the problem. Methods: We analyzed the occurrence of CT in a group of 1226 patients undergoing TLE at a single reference center between June, 2015 and February, 2021. Using standard mechanical devices as first‐line tools, a total of 2092 leads had been extracted. Results: CT occurred in 18 patients (1.47%): due to injury to the wall of the right atrium in 14 patients (1.14%) and other cardiac walls in four patients (0.33%). Younger patient age at first implantation, female gender, high left ventricular ejection fraction (LVEF), lower New York Heart Association class, low Charlson comorbidity index, longer implant duration, and the number of previous procedures related to cardiac implantable electronic devices (CIED) are important patient‐related risk factors for CT. Significant procedure‐related risk factors include the number of extracted leads, extraction of atrial leads and longer dwell time of extracted leads. Intraoperative transoesophageal echocardiography (TEE) provides a lot of information about pulling on various cardiac structures and is able to detect a very early phase of bleeding to the pericardial sac. As a result of implementing best practices guidance in performing extraction procedures and close collaboration with cardiac surgeons that allowed immediate rescue intervention in our series of 18 CT cases, there were no procedure‐related deaths (mortality 0%). Conclusions: The need for rescue surgery due to CT has no influence on clinical and procedural success. Early diagnosed (TEE monitoring) and properly managed CT does not generate any additional risk in short‐ and long‐term follow‐up after TLE. Abstract : The graphical abstract presents the main findings and conclusions … (more)
- Is Part Of:
- Journal of cardiovascular electrophysiology. Volume 33:Number 12(2022)
- Journal:
- Journal of cardiovascular electrophysiology
- Issue:
- Volume 33:Number 12(2022)
- Issue Display:
- Volume 33, Issue 12 (2022)
- Year:
- 2022
- Volume:
- 33
- Issue:
- 12
- Issue Sort Value:
- 2022-0033-0012-0000
- Page Start:
- 2625
- Page End:
- 2639
- Publication Date:
- 2022-09-14
- Subjects:
- cardiac tamponade -- lead extraction complications -- pericardial fluid -- surgery in lead extraction -- transvenous lead extraction
Blood vessels -- Physiology -- Periodicals
Electrophysiology -- Periodicals
Heart -- Physiology -- Periodicals
612.1 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/jce.15668 ↗
- Languages:
- English
- ISSNs:
- 1045-3873
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24755.xml