Long-term integrity of 53 iliac vein stents after catheter-directed thrombolysis. (February 2023)
- Record Type:
- Journal Article
- Title:
- Long-term integrity of 53 iliac vein stents after catheter-directed thrombolysis. (February 2023)
- Main Title:
- Long-term integrity of 53 iliac vein stents after catheter-directed thrombolysis
- Authors:
- Foegh, Pia
Strandberg, Charlotte
Joergensen, Sanne
Myschetzky, Peter S
Klitfod, Lotte
Just, Sven
Hansen, Susanne
Baekgaard, Niels - Abstract:
- Background: Long-term surveillance data on venous stent integrity is sparse. There is limited knowledge on whether duplex ultrasound (DUS) can detect potential stent deformities such as kinking, straightening, and fracture, which may impact long-term patency of the stented veins. Purpose: To assess venous stent integrity after at least five years of follow-up and to establish the efficacy of DUS as surveillance in patients with venous stent. Material and Methods: A total of 45 patients with acute iliac-femoral deep vein thrombosis (DVT) treated with catheter directed thrombolysis (CDT) and stenting >5 years before follow-up. Stents were evaluated with 3D volume low dose non-contrast computed tomography (CT) and DUS for kinking, straightening, stent fracture, and patency. Results from CT scans and DUS were compared to assess the overall agreement between the methods. Results: Median follow-up was 13.2 years (mean = 11.2 years; range = 5.2–15.8 years). 3D CT reconstructions showed normal stent configuration in 47 stents (89%). All intact stents were identified by DUS. In the remaining six stents, 3D CT reconstructions showed compression, tapering, kinking, and minor fracture. DUS recognized all stent complications except the minor fracture. Overall agreement between CT and DUS was 98% (kappa = 0.90). Two cases of stent occlusion were found. Conclusion: The long-term physical resilience of iliac vein stents evaluated with 3D CT in patients treated with CDT for iliofemoral DVTBackground: Long-term surveillance data on venous stent integrity is sparse. There is limited knowledge on whether duplex ultrasound (DUS) can detect potential stent deformities such as kinking, straightening, and fracture, which may impact long-term patency of the stented veins. Purpose: To assess venous stent integrity after at least five years of follow-up and to establish the efficacy of DUS as surveillance in patients with venous stent. Material and Methods: A total of 45 patients with acute iliac-femoral deep vein thrombosis (DVT) treated with catheter directed thrombolysis (CDT) and stenting >5 years before follow-up. Stents were evaluated with 3D volume low dose non-contrast computed tomography (CT) and DUS for kinking, straightening, stent fracture, and patency. Results from CT scans and DUS were compared to assess the overall agreement between the methods. Results: Median follow-up was 13.2 years (mean = 11.2 years; range = 5.2–15.8 years). 3D CT reconstructions showed normal stent configuration in 47 stents (89%). All intact stents were identified by DUS. In the remaining six stents, 3D CT reconstructions showed compression, tapering, kinking, and minor fracture. DUS recognized all stent complications except the minor fracture. Overall agreement between CT and DUS was 98% (kappa = 0.90). Two cases of stent occlusion were found. Conclusion: The long-term physical resilience of iliac vein stents evaluated with 3D CT in patients treated with CDT for iliofemoral DVT was high. Stent deformities were mostly compression, whereas fracture was rarely seen. DUS seems to be sufficient to evaluate venous stent integrity. … (more)
- Is Part Of:
- Acta radiologica. Volume 64:Number 2(2023)
- Journal:
- Acta radiologica
- Issue:
- Volume 64:Number 2(2023)
- Issue Display:
- Volume 64, Issue 2 (2023)
- Year:
- 2023
- Volume:
- 64
- Issue:
- 2
- Issue Sort Value:
- 2023-0064-0002-0000
- Page Start:
- 881
- Page End:
- 886
- Publication Date:
- 2023-02
- Subjects:
- Vascular -- veins -- stenting -- high-resolution computed tomography -- color Doppler ultrasound
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1177/02841851221090118 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24753.xml