First clinical experience with real-time portal imaging-based breath-hold monitoring in tangential breast radiotherapy. (October 2022)
- Record Type:
- Journal Article
- Title:
- First clinical experience with real-time portal imaging-based breath-hold monitoring in tangential breast radiotherapy. (October 2022)
- Main Title:
- First clinical experience with real-time portal imaging-based breath-hold monitoring in tangential breast radiotherapy
- Authors:
- Vasina, Elena N.
Kong, Natalie
Greer, Peter
Baeza Ortega, Jose
Kron, Tomas
Ludbrook, Joanna J.
Thwaites, David
Lehmann, Joerg - Abstract:
- Abstract: Background and purpose: Real-time treatment monitoring with the electronic portal imaging device (EPID) can conceptually provide a more accurate assessment of the quality of deep inspiration breath-hold (DIBH) and patient movement during tangential breast radiotherapy (RT). A system was developed to measure two geometrical parameters, the lung depth (LD) and the irradiated width (named here skin distance, SD), along three user-selected lines in MV EPID images of breast tangents. The purpose of this study was to test the system during tangential breast RT with DIBH. Materials and methods: Measurements of LDs and SDs were carried out in real time. DIBH was guided with a commercial system using a marker block. Results from 17 patients were assessed. Mean midline LDs, , per tangent were compared to the planned mLDs; differences between the largest and smallest observed () per tangent were calculated. Results: For 56% (162/288) of the tangents tested, were outside the tolerance window. All but one patient had at least one fraction showing this behaviour. The largest difference found between an and its planned mLD was −16.9 mm. The accuracy of patient positioning and the quality of marker-block-based DIBH guidance contributed to the differences. Fractions with patient position verification using a single EPID image taken before treatment showed a lower rate (34%), suggesting reassessment of setup procedures. Conclusions: Real-time treatment monitoring of the internalAbstract: Background and purpose: Real-time treatment monitoring with the electronic portal imaging device (EPID) can conceptually provide a more accurate assessment of the quality of deep inspiration breath-hold (DIBH) and patient movement during tangential breast radiotherapy (RT). A system was developed to measure two geometrical parameters, the lung depth (LD) and the irradiated width (named here skin distance, SD), along three user-selected lines in MV EPID images of breast tangents. The purpose of this study was to test the system during tangential breast RT with DIBH. Materials and methods: Measurements of LDs and SDs were carried out in real time. DIBH was guided with a commercial system using a marker block. Results from 17 patients were assessed. Mean midline LDs, , per tangent were compared to the planned mLDs; differences between the largest and smallest observed () per tangent were calculated. Results: For 56% (162/288) of the tangents tested, were outside the tolerance window. All but one patient had at least one fraction showing this behaviour. The largest difference found between an and its planned mLD was −16.9 mm. The accuracy of patient positioning and the quality of marker-block-based DIBH guidance contributed to the differences. Fractions with patient position verification using a single EPID image taken before treatment showed a lower rate (34%), suggesting reassessment of setup procedures. Conclusions: Real-time treatment monitoring of the internal anatomy during DIBH delivery of tangential breast RT is feasible and useful. The new system requires no additional radiation for the patient. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 24(2022)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 24(2022)
- Issue Display:
- Volume 24, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 24
- Issue:
- 2022
- Issue Sort Value:
- 2022-0024-2022-0000
- Page Start:
- 1
- Page End:
- 6
- Publication Date:
- 2022-10
- Subjects:
- Real-time monitoring -- Electronic portal imaging device -- EPID -- Tangential breast radiotherapy -- Internal anatomy -- Deep inspiration breath-hold -- DIBH -- Lung depth -- Intra-fraction motion
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2022.08.002 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24702.xml