Characterization of a miniaturized scintillator detector for time-resolved treatment monitoring in HDR-brachytherapy. (21st December 2022)
- Record Type:
- Journal Article
- Title:
- Characterization of a miniaturized scintillator detector for time-resolved treatment monitoring in HDR-brachytherapy. (21st December 2022)
- Main Title:
- Characterization of a miniaturized scintillator detector for time-resolved treatment monitoring in HDR-brachytherapy
- Authors:
- Gonod, Mathieu
Suarez, Miguel Angel
Chacon Avila, Carlos
Karakhanyan, Vage
Eustache, Clément
Crouzilles, Julien
Laskri, Samir
Vinchant, Jean-François
Aubignac, Léone
Grosjean, Thierry - Abstract:
- Abstract: Purpose. HDR brachytherapy combines steep dose gradients in space and time, thereby requiring detectors of high spatial and temporal resolution to perform accurate treatment monitoring. We demonstrate a miniaturized fiber-integrated scintillator detector (MSD) of unmatched compactness which fulfills these conditions. Methods. The MSD consists of a 0.28 mm large and 0.43 mm long detection cell (Gd2 O2 S:Tb) coupled to a 110 micron outer diameter silica optical fiber. The fiber probe is tested in a phantom using a MicroSelectron 9.1 Ci Ir-192 HDR afterloader. The detection signal is acquired at a rate of 0.08 s with a standard sCMOS camera coupled to a chromatic filter (to cancel spurious Cerenkov signal). The dwell position and time monitoring are analyzed over prostate treatment sequences with dwell times spanning from 0.1 to 11 s. The dose rate at the probe position is both evaluated from a direct measurement and by reconstruction from the measured dwell position using the AAPM TG-43 formalism. Results. A total number of 1384 dwell positions are analyzed. In average, the measured dwell positions differ by 0.023 ± 0.077 mm from planned values over a 6–54 mm source-probe distance range. The standard deviation of the measured dwell positions is below 0.8 mm. 94% of the 966 dwell positions occurring at a source-probe inter-catheter spacing below 20 mm are successfully identified, with a 100% detection rate for dwell times exceeding 0.5 s. The average deviation to theAbstract: Purpose. HDR brachytherapy combines steep dose gradients in space and time, thereby requiring detectors of high spatial and temporal resolution to perform accurate treatment monitoring. We demonstrate a miniaturized fiber-integrated scintillator detector (MSD) of unmatched compactness which fulfills these conditions. Methods. The MSD consists of a 0.28 mm large and 0.43 mm long detection cell (Gd2 O2 S:Tb) coupled to a 110 micron outer diameter silica optical fiber. The fiber probe is tested in a phantom using a MicroSelectron 9.1 Ci Ir-192 HDR afterloader. The detection signal is acquired at a rate of 0.08 s with a standard sCMOS camera coupled to a chromatic filter (to cancel spurious Cerenkov signal). The dwell position and time monitoring are analyzed over prostate treatment sequences with dwell times spanning from 0.1 to 11 s. The dose rate at the probe position is both evaluated from a direct measurement and by reconstruction from the measured dwell position using the AAPM TG-43 formalism. Results. A total number of 1384 dwell positions are analyzed. In average, the measured dwell positions differ by 0.023 ± 0.077 mm from planned values over a 6–54 mm source-probe distance range. The standard deviation of the measured dwell positions is below 0.8 mm. 94% of the 966 dwell positions occurring at a source-probe inter-catheter spacing below 20 mm are successfully identified, with a 100% detection rate for dwell times exceeding 0.5 s. The average deviation to the planned dwell times is of 0.005 ± 0.060 s. The instant dose retrieval from dwell position monitoring leads to a relative mismatch to planned values of 0.14% ± 0.7%. Conclusion. A miniaturized Gd2O2 S:Tb detector coupled to a standard sCMOS camera can be used for time-resolved treatment monitoring in HDR Brachytherapy. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 67:Number 24(2022)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 67:Number 24(2022)
- Issue Display:
- Volume 67, Issue 24 (2022)
- Year:
- 2022
- Volume:
- 67
- Issue:
- 24
- Issue Sort Value:
- 2022-0067-0024-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-12-21
- Subjects:
- HDR brachytherapy -- in vivo dosimetry -- miniaturized scintillator detector -- source tracking -- dwell position monitoring -- dwell time verification -- dose rate monitoring
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/ac9a9b ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24690.xml