Characterization of intragastric pressure waveform in endoscopic pressure study integrated system: Novel diagnostic device for gastroesophageal reflux disease. Issue 5 (8th December 2020)
- Record Type:
- Journal Article
- Title:
- Characterization of intragastric pressure waveform in endoscopic pressure study integrated system: Novel diagnostic device for gastroesophageal reflux disease. Issue 5 (8th December 2020)
- Main Title:
- Characterization of intragastric pressure waveform in endoscopic pressure study integrated system: Novel diagnostic device for gastroesophageal reflux disease
- Authors:
- Shimamura, Yuto
Inoue, Haruhiro
Rodriguez de Santiago, Enrique
Abad, Mary Raina Angeli
Fujiyoshi, Yusuke
Toshimori, Akiko
Tanabe, Mayo
Sumi, Kazuya
Iwaya, Yugo
Ikeda, Haruo
Onimaru, Manabu - Abstract:
- Abstract : Objectives: Endoscopic pressure study integrated system (EPSIS) is a novel diagnostic tool for gastroesophageal reflux disease (GERD). EPSIS has been developed to evaluate lower esophageal sphincter function by monitoring the intragastric pressure (IGP) while insufflating the stomach during gastroscopy. Based on previous data, EPSIS could diagnose GERD with good accuracy by assessing IGP waveform pattern. This study aimed to further characterize the waveform to improve the diagnostic yield of EPSIS. Methods: We conducted a retrospective analysis of patients with typical GERD symptoms who underwent both EPSIS and 24‐h impedance‐pH monitoring (MII‐pH) at a single tertiary referral center from October 2018 to May 2020. EPSIS was performed by using a through‐the‐scope catheter connected to the pressure measuring system (TR‐W550, TR‐TeH08, AP‐C35; Keyence, Osaka, Japan) to monitor IGP. Abnormal acid reflux was defined as acid exposure time (AET) over 6.0%. Pressure waveform was characterized as follows: (i) Basal IGP, (ii) Maximum IGP, (iii) Pressure difference, (iv) Gradient of the waveform. Results: A total of 57 patients with GERD symptoms were analyzed. Twenty‐one patients presented abnormal AET on MII‐pH. Among EPSIS parameters, pressure difference during insufflation correlated with AET (ρ = −0.66, P < 0.01) and showed the best diagnostic accuracy for AET with the cutoff value of 4.7 mmHg (area under the curve [AUC], 0.87). The gradient of EPSIS waveform alsoAbstract : Objectives: Endoscopic pressure study integrated system (EPSIS) is a novel diagnostic tool for gastroesophageal reflux disease (GERD). EPSIS has been developed to evaluate lower esophageal sphincter function by monitoring the intragastric pressure (IGP) while insufflating the stomach during gastroscopy. Based on previous data, EPSIS could diagnose GERD with good accuracy by assessing IGP waveform pattern. This study aimed to further characterize the waveform to improve the diagnostic yield of EPSIS. Methods: We conducted a retrospective analysis of patients with typical GERD symptoms who underwent both EPSIS and 24‐h impedance‐pH monitoring (MII‐pH) at a single tertiary referral center from October 2018 to May 2020. EPSIS was performed by using a through‐the‐scope catheter connected to the pressure measuring system (TR‐W550, TR‐TeH08, AP‐C35; Keyence, Osaka, Japan) to monitor IGP. Abnormal acid reflux was defined as acid exposure time (AET) over 6.0%. Pressure waveform was characterized as follows: (i) Basal IGP, (ii) Maximum IGP, (iii) Pressure difference, (iv) Gradient of the waveform. Results: A total of 57 patients with GERD symptoms were analyzed. Twenty‐one patients presented abnormal AET on MII‐pH. Among EPSIS parameters, pressure difference during insufflation correlated with AET (ρ = −0.66, P < 0.01) and showed the best diagnostic accuracy for AET with the cutoff value of 4.7 mmHg (area under the curve [AUC], 0.87). The gradient of EPSIS waveform also revealed good diagnostic accuracy for abnormal AET with the cutoff value of 0.07 mmHg/s (AUC, 0.81). Conclusions: Endoscopic pressure study integrated system waveform parameters, especially pressure difference, presented high diagnostic accuracy for the presence of abnormal acid reflux. … (more)
- Is Part Of:
- Digestive endoscopy. Volume 33:Issue 5(2021)
- Journal:
- Digestive endoscopy
- Issue:
- Volume 33:Issue 5(2021)
- Issue Display:
- Volume 33, Issue 5 (2021)
- Year:
- 2021
- Volume:
- 33
- Issue:
- 5
- Issue Sort Value:
- 2021-0033-0005-0000
- Page Start:
- 780
- Page End:
- 787
- Publication Date:
- 2020-12-08
- Subjects:
- endoscopic diagnosis -- endoscopy -- esophagus -- gastroesophageal reflux disease
Digestive organs -- Diseases -- Periodicals
Digestive organs -- Diseases -- Diagnosis -- Periodicals
Endoscopy -- Periodicals
Digestive System Diseases -- diagnosis -- Periodicals
Digestive System Diseases -- therapy -- Periodicals
Endoscopy -- Periodicals
616.3 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/den.13867 ↗
- Languages:
- English
- ISSNs:
- 0915-5635
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3588.346200
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24659.xml