Flow generators for helmet CPAP: Which to prefer? A bench study. (February 2023)
- Record Type:
- Journal Article
- Title:
- Flow generators for helmet CPAP: Which to prefer? A bench study. (February 2023)
- Main Title:
- Flow generators for helmet CPAP: Which to prefer? A bench study
- Authors:
- Privitera, Daniele
Capsoni, Nicolò
Zadek, Francesco
Mazzone, Annamaria
Bellone, Andrea
Fumagalli, Roberto
Langer, Thomas - Abstract:
- Abstract: Objective: To assess the different effect of filters' application during helmet-CPAP delivered with three different flow generators on the delivered fresh gas flow, FiO2, and the noise level inside and outside the helmet. Methods: In a bench study, three flow generators (air-oxygen blender, turbine ventilator and Venturi system) were used to generate two different gas flows (60 L/min and 80 L/min), with a fixed FiO2 at 0.6, to perform a helmet-CPAP on a manikin. Three different fixed PEEP valves (7.5, 10, and 12.5 cmH2 O) were applied at the expiratory port. Gas flow, FiO2 and noise were recorded for each Flow-generator/Flow/PEEP combination, first without filter interposition and then after positioning a heat and moister exchanger filter (HMEF) at the helmet inlet port. Results: The application of the HMEF lead to a significant difference in the flow variation among the three flow generators (p < 0.001). Compared to baseline, the highest flow reduction was observed with the VENTURI (−13.4 ± 1.2 %, p < 0.001), a slight increase with the BLENDER (1.2 ± 0.5 %, p < 0.001), whereas no difference was recorded with the TURBINE (0.1 ± 0.6 %, p = 0.12). After HMEF was interposed, a significant FiO2 variation was observed only with VENTURI (11.3 ± 1.8 %, p < 0.001). As for the noise, the TURBINE was the least noisy system, both with and without the filter interposition. Conclusions: Flow generators used to deliver helmet-CPAP have different characteristics and responses toAbstract: Objective: To assess the different effect of filters' application during helmet-CPAP delivered with three different flow generators on the delivered fresh gas flow, FiO2, and the noise level inside and outside the helmet. Methods: In a bench study, three flow generators (air-oxygen blender, turbine ventilator and Venturi system) were used to generate two different gas flows (60 L/min and 80 L/min), with a fixed FiO2 at 0.6, to perform a helmet-CPAP on a manikin. Three different fixed PEEP valves (7.5, 10, and 12.5 cmH2 O) were applied at the expiratory port. Gas flow, FiO2 and noise were recorded for each Flow-generator/Flow/PEEP combination, first without filter interposition and then after positioning a heat and moister exchanger filter (HMEF) at the helmet inlet port. Results: The application of the HMEF lead to a significant difference in the flow variation among the three flow generators (p < 0.001). Compared to baseline, the highest flow reduction was observed with the VENTURI (−13.4 ± 1.2 %, p < 0.001), a slight increase with the BLENDER (1.2 ± 0.5 %, p < 0.001), whereas no difference was recorded with the TURBINE (0.1 ± 0.6 %, p = 0.12). After HMEF was interposed, a significant FiO2 variation was observed only with VENTURI (11.3 ± 1.8 %, p < 0.001). As for the noise, the TURBINE was the least noisy system, both with and without the filter interposition. Conclusions: Flow generators used to deliver helmet-CPAP have different characteristics and responses to HMEF interposition. Users should be aware of the effects on FiO2 and flow of the different devices in order to make a precise setup of the circuit. … (more)
- Is Part Of:
- Intensive and critical care nursing. Volume 74(2023)
- Journal:
- Intensive and critical care nursing
- Issue:
- Volume 74(2023)
- Issue Display:
- Volume 74, Issue 2023 (2023)
- Year:
- 2023
- Volume:
- 74
- Issue:
- 2023
- Issue Sort Value:
- 2023-0074-2023-0000
- Page Start:
- Page End:
- Publication Date:
- 2023-02
- Subjects:
- Continuous positive airway pressure -- Helmet -- Non-invasive ventilation -- Respiratory insufficiency -- Noise -- Filter -- Emergency department
Intensive care nursing -- Periodicals
Critical Illness -- nursing -- Periodicals
Intensive Care -- Periodicals
Nursing Care -- Periodicals
Intensive care nursing
Periodicals
Electronic journals
616.028 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09643397 ↗
http://www.harcourt-international.com/journal ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.iccn.2022.103344 ↗
- Languages:
- English
- ISSNs:
- 0964-3397
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4531.836000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24630.xml