Proof of principle of a novel co‐pulsating intra‐ventricular membrane pump. Issue 12 (22nd July 2020)
- Record Type:
- Journal Article
- Title:
- Proof of principle of a novel co‐pulsating intra‐ventricular membrane pump. Issue 12 (22nd July 2020)
- Main Title:
- Proof of principle of a novel co‐pulsating intra‐ventricular membrane pump
- Authors:
- van Dort, Daniël I. M.
Thannhauser, Jos
Gommans, Frank D. H.
Ten Cate, Tim J.
Duncker, Dirk J.
Suryapranata, Harry
Morshuis, Wim J.
Geuzebroek, Guillaume S. C. - Abstract:
- Abstract: In this proof of principle study, we investigated the effectiveness and safety of hemodynamic support with the Intra‐Ventricular Membrane Pump (IVMP). The IVMP was implanted into the apex of the left ventricle. Hemodynamic assessment was performed in six ex vivo beating porcine hearts (PhysioHeart platform). The cardiac output (CO), mean arterial pressure (MAP), coronary flow (CF) and pulse pressure (PP) were obtained before and during IVMP support and reported as means ± standard deviations. In two additional visualization experiments, the integrity of the mitral valve was assessed during IVMP support. We found a significant increase of the CO (+1.4 ± 0.2 L/min, P < .001), MAP (+13 ± 6 mm Hg, P = .008), CF (+0.23 ± 0.1 L/min, P = .004), and PP (+15 ± 4 mm Hg, P = .002) during IVMP support, when compared to baseline. No interference of the IVMP with mitral valve function was observed. An increase of premature ventricular complexes (PVC) was observed during support with the IVMP (mean PVC‐burden 4.3% vs. 0.7% at baseline), negatively influencing hemodynamic parameters. The IVMP is able to significantly improve hemodynamic parameters in a co‐pulsatile fashion, without hampering the function of the mitral valve. These findings provide a basis for future development of a catheter‐based IVMP.
- Is Part Of:
- Artificial organs. Volume 44:Issue 12(2020)
- Journal:
- Artificial organs
- Issue:
- Volume 44:Issue 12(2020)
- Issue Display:
- Volume 44, Issue 12 (2020)
- Year:
- 2020
- Volume:
- 44
- Issue:
- 12
- Issue Sort Value:
- 2020-0044-0012-0000
- Page Start:
- 1267
- Page End:
- 1275
- Publication Date:
- 2020-07-22
- Subjects:
- cardiogenic shock -- ex vivo beating heart -- mechanical circulatory support -- pulsatile flow
Artificial organs -- Periodicals
617.956 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1525-1594 ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=aor ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1111/aor.13757 ↗
- Languages:
- English
- ISSNs:
- 0160-564X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1735.052000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24588.xml