Analysis of Polyethylene-Related Revisions After Total Ankle Replacements Reported in US Food and Drug Administration Medical Device Adverse Event Database. (January 2023)
- Record Type:
- Journal Article
- Title:
- Analysis of Polyethylene-Related Revisions After Total Ankle Replacements Reported in US Food and Drug Administration Medical Device Adverse Event Database. (January 2023)
- Main Title:
- Analysis of Polyethylene-Related Revisions After Total Ankle Replacements Reported in US Food and Drug Administration Medical Device Adverse Event Database
- Authors:
- Jiang, Hongying
Wu, Leihong
Randsborg, Per-Henrik
Houck, Jennifer
Sun, Limin
Marine, Marissa
Chow, Megan
Peluso, Joseph
Peat, Raquel - Abstract:
- Background: There are 2 general types of total ankle replacement (TAR) designs with respect to the polyethylene insert, mobile-bearing (MB) and fixed-bearing (FB) TARs. The aim of this study is to compare polyethylene-related adverse events (AEs), particularly revisions, reported for MB TARs and FB TARs using the US Food and Drug Administration's (FDA's) Manufacturer and User Facility Device Experience (MAUDE) database. Methods: A text mining method was applied to the medical device reporting (MDR) in the MAUDE database from 1991 to 2020, followed by manual reviews to identify, characterize, and describe all polyethylene-related AEs, including revisions, reported for MB and FB TARs. Results: We found 1841 MDRs for MB (STAR Ankle only) and 1273 MDRs for 40+ FB TARs approved/cleared by the FDA. For the MB design, 33% (606/1841) of the AEs reported related to the polyethylene component, compared to 24% (291/1273) of the AEs reported for FB designs. Polyethylene fractures were reported in 11.3% (208/1841) for the MB designs compared to 0.2% (2/1273) for the FB designs. Half of the polyethylene-related revisions occurred within an average of 4.1 years after implantation for the MB design compared within an average of 5.2 years for FB designs. Conclusion: Analysis of this database revealed a higher proportion of reported polyethylene fractures and greater need for earlier revisions for polyethylene-related issues with use of the primary MB design in the database as compared withBackground: There are 2 general types of total ankle replacement (TAR) designs with respect to the polyethylene insert, mobile-bearing (MB) and fixed-bearing (FB) TARs. The aim of this study is to compare polyethylene-related adverse events (AEs), particularly revisions, reported for MB TARs and FB TARs using the US Food and Drug Administration's (FDA's) Manufacturer and User Facility Device Experience (MAUDE) database. Methods: A text mining method was applied to the medical device reporting (MDR) in the MAUDE database from 1991 to 2020, followed by manual reviews to identify, characterize, and describe all polyethylene-related AEs, including revisions, reported for MB and FB TARs. Results: We found 1841 MDRs for MB (STAR Ankle only) and 1273 MDRs for 40+ FB TARs approved/cleared by the FDA. For the MB design, 33% (606/1841) of the AEs reported related to the polyethylene component, compared to 24% (291/1273) of the AEs reported for FB designs. Polyethylene fractures were reported in 11.3% (208/1841) for the MB designs compared to 0.2% (2/1273) for the FB designs. Half of the polyethylene-related revisions occurred within an average of 4.1 years after implantation for the MB design compared within an average of 5.2 years for FB designs. Conclusion: Analysis of this database revealed a higher proportion of reported polyethylene fractures and greater need for earlier revisions for polyethylene-related issues with use of the primary MB design in the database as compared with FB TAR designs. Further study of device-related complications with more recent designs for both MB and FB ankle replacement components are needed to improve the outcomes of total ankle replacement. Level of Evidence: Level III, retrospective comparative study. … (more)
- Is Part Of:
- Foot & ankle international. Volume 44:Number 1(2023)
- Journal:
- Foot & ankle international
- Issue:
- Volume 44:Number 1(2023)
- Issue Display:
- Volume 44, Issue 1 (2023)
- Year:
- 2023
- Volume:
- 44
- Issue:
- 1
- Issue Sort Value:
- 2023-0044-0001-0000
- Page Start:
- 13
- Page End:
- 20
- Publication Date:
- 2023-01
- Subjects:
- total ankle replacement -- TAR -- adverse event -- revision -- polyethylene-related adverse events -- polyethylene fractures -- polyethylene wear -- FDA -- FDA Manufacturer and User Facility Device Experience -- MAUDE -- MDR -- ultrahigh molecular weight polyethylene -- UHMWPE
Foot -- Abnormalities -- Periodicals
Ankle -- Abnormalities -- Periodicals
Orthopedics -- Periodicals
617.585 - Journal URLs:
- http://fai.sagepub.com/ ↗
http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&NEWS=n&PAGE=toc&D=ovft&AN=00041550-000000000-00000 ↗
http://www.sagepublications.com/ ↗
http://207.158.206.46/medical/FAI_body.htm ↗
http://www.datatrace.com/medical/FAI_online.htm ↗ - DOI:
- 10.1177/10711007221134284 ↗
- Languages:
- English
- ISSNs:
- 1071-1007
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24551.xml