Effects of Simulated Adverse Environmental Conditions Related to Actual Conditions at Health Promoting Hospitals on the Performance of Blood Glucose Testing by Glucose Meters. (January 2023)
- Record Type:
- Journal Article
- Title:
- Effects of Simulated Adverse Environmental Conditions Related to Actual Conditions at Health Promoting Hospitals on the Performance of Blood Glucose Testing by Glucose Meters. (January 2023)
- Main Title:
- Effects of Simulated Adverse Environmental Conditions Related to Actual Conditions at Health Promoting Hospitals on the Performance of Blood Glucose Testing by Glucose Meters
- Authors:
- Apiratmateekul, Napaporn
Duanginta, Wanutchaya
Phetree, Methawadee
Kongros, Kunchit
Treebuphachatsakul, Wanvisa - Abstract:
- Background: Annual peaks in temperature and humidity exceed manufacturers' specifications for blood glucose test strip storage in Thailand. Health Promoting Hospitals (HPH) do not always provide the same level of health services that hospitals do since they often only turn on air-conditioning units during working hours. Methods: The blood glucose testing performance of 4 glucose meters were investigated for short and long terms stress at 5 simulated conditions. Temperature and relative humidity (RH) at 5 HPHs in different regions of Thailand were monitored for 9 weeks during April to July 2019. The use of air conditioning in storage rooms for storing test strips was surveyed at 131 HPHs using questionnaires. Results: Median-paired differences of glucose measurements with stressed test strips in 5 simulated conditions significantly differed ( P < 0.05) both in the short term (3 days) and in the long term (30 days) with 4 glucose meters when compared to unstressed test strips. The average temperature of all HPHs exceeded 30°C (86°F). The average RH was 84%. There was only one HPH that occasionally turned on its air conditioners. Most HPHs kept both opened and unopened vials of test strips in rooms without air conditioners. Further, 21.4-32.0% of HPHs kept test strips at room with air conditioners. Conclusions: This study provides evidence for poor performance of blood glucose testing by glucose meters that are affected by adverse environmental conditions. The environmentalBackground: Annual peaks in temperature and humidity exceed manufacturers' specifications for blood glucose test strip storage in Thailand. Health Promoting Hospitals (HPH) do not always provide the same level of health services that hospitals do since they often only turn on air-conditioning units during working hours. Methods: The blood glucose testing performance of 4 glucose meters were investigated for short and long terms stress at 5 simulated conditions. Temperature and relative humidity (RH) at 5 HPHs in different regions of Thailand were monitored for 9 weeks during April to July 2019. The use of air conditioning in storage rooms for storing test strips was surveyed at 131 HPHs using questionnaires. Results: Median-paired differences of glucose measurements with stressed test strips in 5 simulated conditions significantly differed ( P < 0.05) both in the short term (3 days) and in the long term (30 days) with 4 glucose meters when compared to unstressed test strips. The average temperature of all HPHs exceeded 30°C (86°F). The average RH was 84%. There was only one HPH that occasionally turned on its air conditioners. Most HPHs kept both opened and unopened vials of test strips in rooms without air conditioners. Further, 21.4-32.0% of HPHs kept test strips at room with air conditioners. Conclusions: This study provides evidence for poor performance of blood glucose testing by glucose meters that are affected by adverse environmental conditions. The environmental for test strips storage at HPHs should be considered to prevent analytical errors of glucose measurement. … (more)
- Is Part Of:
- Journal of diabetes science and technology. Volume 17:Number 1(2023)
- Journal:
- Journal of diabetes science and technology
- Issue:
- Volume 17:Number 1(2023)
- Issue Display:
- Volume 17, Issue 1 (2023)
- Year:
- 2023
- Volume:
- 17
- Issue:
- 1
- Issue Sort Value:
- 2023-0017-0001-0000
- Page Start:
- 125
- Page End:
- 132
- Publication Date:
- 2023-01
- Subjects:
- temperature stress test -- relative humidity -- accuracy of blood glucose by glucose meter -- health promoting hospitals
Diabetes -- Periodicals
Medical technology -- Periodicals
Diabetes Mellitus -- Periodicals
616.462005 - Journal URLs:
- http://ejournals.ebsco.com/direct.asp?JournalID=712321 ↗
http://www.jodsat.org/about.html ↗
http://online.sagepub.com/ ↗ - DOI:
- 10.1177/19322968211042343 ↗
- Languages:
- English
- ISSNs:
- 1932-2968
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24545.xml