Post-Market Surveillance Assessment of the Clinical Accuracy of a Blood Glucose Monitoring System with an Improved Algorithm for Enhanced Product Performance. (January 2023)
- Record Type:
- Journal Article
- Title:
- Post-Market Surveillance Assessment of the Clinical Accuracy of a Blood Glucose Monitoring System with an Improved Algorithm for Enhanced Product Performance. (January 2023)
- Main Title:
- Post-Market Surveillance Assessment of the Clinical Accuracy of a Blood Glucose Monitoring System with an Improved Algorithm for Enhanced Product Performance
- Authors:
- Setford, Steven
Liu, Zuifang
McColl, David
Phillips, Stuart
Cameron, Hilary
Grady, Mike - Abstract:
- Background: On-going manufacturer-led post-market surveillance (PMS), assessing the clinical accuracy of blood glucose monitoring (BGM) systems, is critical to substantiate the performance of such products for people with diabetes. Materials and Methods: Batches of Verio test-strip product were randomly and routinely selected over the period from launch of an improved-algorithm product to reporting date and sent to 3 clinic sites for clinician-led accuracy assessment. Accuracy is reported as per recently adopted FDA guidance for BGM systems, EN ISO 15197:2015 and MARD/MAD (Mean absolute relative difference/Mean absolute difference). Results: Thirty-three individual test-strip batches were evaluated corresponding to 506 unique donors. Accuracy performance - FDA: 98.9% of values within ±15% of comparator; ISO: 99.0% within ±15 mg/dL or ±15% at <100 mg/dL (<5.55 mmol/L) or ≥100 mg/dL (≥5.55 mmol/L) glucose, respectively. Overall MARD was 4.19% with a MARD range of 3.54%-5.73% across all test strip batches. Conclusions: This post-market surveillance program demonstrates the new BGM system consistently meets measures of clinical accuracy specified by regulators. This program supports a growing demand by regulators for real-world evidence demonstrating consistent in-market product efficacy as opposed to the current largely passive approach that relies on assessment of reports filed by device users.
- Is Part Of:
- Journal of diabetes science and technology. Volume 17:Number 1(2023)
- Journal:
- Journal of diabetes science and technology
- Issue:
- Volume 17:Number 1(2023)
- Issue Display:
- Volume 17, Issue 1 (2023)
- Year:
- 2023
- Volume:
- 17
- Issue:
- 1
- Issue Sort Value:
- 2023-0017-0001-0000
- Page Start:
- 133
- Page End:
- 140
- Publication Date:
- 2023-01
- Subjects:
- post-market surveillance (PMS) -- accuracy -- MARD -- blood glucose monitoring (BGM) -- regulations
Diabetes -- Periodicals
Medical technology -- Periodicals
Diabetes Mellitus -- Periodicals
616.462005 - Journal URLs:
- http://ejournals.ebsco.com/direct.asp?JournalID=712321 ↗
http://www.jodsat.org/about.html ↗
http://online.sagepub.com/ ↗ - DOI:
- 10.1177/19322968211039465 ↗
- Languages:
- English
- ISSNs:
- 1932-2968
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24545.xml