Verification of the Bühlmann fPELA turbo faecal elastase assay on the Binding Site Optilite benchtop analyser. (January 2023)
- Record Type:
- Journal Article
- Title:
- Verification of the Bühlmann fPELA turbo faecal elastase assay on the Binding Site Optilite benchtop analyser. (January 2023)
- Main Title:
- Verification of the Bühlmann fPELA turbo faecal elastase assay on the Binding Site Optilite benchtop analyser
- Authors:
- Handley, Simon A
Dote, Nikko P
Wanandy, Troy
Prentice, Louise - Abstract:
- Background: Pancreatic elastase-1 (PE1) can be measured to assess exocrine activity of the pancreas. A semi-automated particle-enhanced, open-channel turbidimetric immunoassay has been introduced by Bühlmann (fPELA turbo, Bühlmann Laboratories AG, Schoenenbuch, Switzerland). Published evaluation data is lacking. We therefore verified performance of the assay on the Binding Site Optilite benchtop analyser and undertook a sample comparison with the DiaSorin PE1 assay on the Liaison. Methods: Accuracy, imprecision, lower limit of quantitation (LLoQ) and linearity of the Bühlmann fPELA turbo assay on the Binding Site Optilite analyser was ascertained. Comparison with the DiaSorin Liaison PE1 assay was also undertaken. Difference between assays was evaluated using the Wilcoxon signed-rank test and method comparison was undertaken using Spearman's rank correlation (rs), Bland–Altman and Passing–Bablok regression analyses. Results: The fPELA turbo assay was linear between 5 and 2500 μg/g. The LLoQ was 5 µg/g. Intra- and inter-assay imprecision was <6%. There was a good agreement (rs = 0.92) and no significant bias (5.8 µg/g, P = 0.29) present between the Bühlmann fPELA turbo and DiaSorin PE1 assays. Conclusion: The Bühlmann fPELA turbo assay performs well on the Binding Site Optilite analyser. Faecal elastase results are commutable between with Bühlmann fPELA turbo and DiaSorin Liaison PE1 assays.
- Is Part Of:
- Annals of clinical biochemistry. Volume 60:Number 1(2023)
- Journal:
- Annals of clinical biochemistry
- Issue:
- Volume 60:Number 1(2023)
- Issue Display:
- Volume 60, Issue 1 (2023)
- Year:
- 2023
- Volume:
- 60
- Issue:
- 1
- Issue Sort Value:
- 2023-0060-0001-0000
- Page Start:
- 68
- Page End:
- 71
- Publication Date:
- 2023-01
- Subjects:
- Elastase -- immunoassay -- pancreatic insufficiency
Clinical chemistry -- Periodicals
Clinical biochemistry -- Periodicals
616.075 - Journal URLs:
- http://web.ebscohost.com/ehost/detail?sid=810a7788-77dd-439f-9630-ad7f5b199fd3%40sessionmgr4&vid=1&hid=14&bdata=JnNpdGU9ZWhvc3QtbGl2ZSZzY29wZT1zaXRl#db=mnh&jid=0324055 ↗
http://acb.rsmjournals.com ↗
http://www.usc.edu/hsc/nml/e-resources/info/annclib.html ↗
http://www.uk.sagepub.com/home.nav ↗
http://www.ingentaconnect.com/content/rsm/acb ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1177/00045632221138025 ↗
- Languages:
- English
- ISSNs:
- 0004-5632
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24542.xml