Study of thrombin generation with St Genesia to evaluate xaban pharmacodynamics: Analytical performances over 18 months. (28th December 2020)
- Record Type:
- Journal Article
- Title:
- Study of thrombin generation with St Genesia to evaluate xaban pharmacodynamics: Analytical performances over 18 months. (28th December 2020)
- Main Title:
- Study of thrombin generation with St Genesia to evaluate xaban pharmacodynamics: Analytical performances over 18 months
- Authors:
- Foulon‐Pinto, Geoffrey
Jourdi, Georges
Perrin, Julien
Abdoul, Johan
Paris, Guillaume
Gouin‐Thibault, Isabelle
Curis, Emmanuel
Lecompte, Thomas
Siguret, Virginie - Abstract:
- Abstract: Introduction: ST Genesia is a new automated system enabling quantitative standardized evaluation of thrombin generation (TG), for example, in patients receiving anti‐Xa direct inhibitors (xabans). Data on its analytical performances are scarce. Methods: Over an 18‐month period, repeatability, reproducibility, and accuracy were assessed using STG‐ThromboScreen (without or with thrombomodulin) or STG‐DrugScreen reagents (corresponding to intermediate/high tissue‐factor concentration, respectively), and controls. Furthermore, reproducibility was assessed using commercialized lyophilized and frozen normal pooled plasmas. Rivaroxaban and apixaban impacts on TG parameters were assessed using spiking experiments. Finally, a comparison with the Calibrated Automated Thrombogram method (CAT) (PPP reagent) was performed using plasma from healthy volunteers enrolled in the DRIVING‐studyNCT 01627665) before and after rivaroxaban intake. Results: For all dedicated quality control (QC) levels, inter‐series coefficients of variations (CV) were <7% for temporal TG parameters, peak height (PH), and endogenous thrombin potential (ETP), whether results were normalized with a dedicated reference plasma STG‐RefPlasma or not. Noteworthy, STG‐RefPlasma used for normalization displayed substantially high PH and ETP. Mean biases between the observed and manufacturer's assigned QC values were mostly <7%. Both rivaroxaban/apixaban plasma concentrations were significantly associated with TGAbstract: Introduction: ST Genesia is a new automated system enabling quantitative standardized evaluation of thrombin generation (TG), for example, in patients receiving anti‐Xa direct inhibitors (xabans). Data on its analytical performances are scarce. Methods: Over an 18‐month period, repeatability, reproducibility, and accuracy were assessed using STG‐ThromboScreen (without or with thrombomodulin) or STG‐DrugScreen reagents (corresponding to intermediate/high tissue‐factor concentration, respectively), and controls. Furthermore, reproducibility was assessed using commercialized lyophilized and frozen normal pooled plasmas. Rivaroxaban and apixaban impacts on TG parameters were assessed using spiking experiments. Finally, a comparison with the Calibrated Automated Thrombogram method (CAT) (PPP reagent) was performed using plasma from healthy volunteers enrolled in the DRIVING‐studyNCT 01627665) before and after rivaroxaban intake. Results: For all dedicated quality control (QC) levels, inter‐series coefficients of variations (CV) were <7% for temporal TG parameters, peak height (PH), and endogenous thrombin potential (ETP), whether results were normalized with a dedicated reference plasma STG‐RefPlasma or not. Noteworthy, STG‐RefPlasma used for normalization displayed substantially high PH and ETP. Mean biases between the observed and manufacturer's assigned QC values were mostly <7%. Both rivaroxaban/apixaban plasma concentrations were significantly associated with TG parameters. Finally, Bland‐Altman plots showed a good agreement between ST Genesia‐STG‐ThromboScreen and CAT method within the explored range of values, although biases could be observed (PH: 16.4 ± 13.2%, ETP: 17.8 ± 11.9%). Conclusion: ST Genesia ® enables the reliable measurement of TG parameters in both in vitro and ex vivo xaban plasma samples using either STG‐ThromboScreen or STG‐DrugScreen according to xaban concentrations. The use of reference plasma, despite not completely reflecting a normal pooled plasma behavior, likely improves standardization and inter‐laboratory comparisons. … (more)
- Is Part Of:
- International journal of laboratory hematology. Volume 43:Number 4(2021)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 43:Number 4(2021)
- Issue Display:
- Volume 43, Issue 4 (2021)
- Year:
- 2021
- Volume:
- 43
- Issue:
- 4
- Issue Sort Value:
- 2021-0043-0004-0000
- Page Start:
- 821
- Page End:
- 830
- Publication Date:
- 2020-12-28
- Subjects:
- anticoagulants -- coagulation -- quality controls -- thrombin generation -- xaban
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.13443 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24477.xml