Cross-camera comparison of ROI-based semi-quantitative 123I-IBZM SPECT data in healthy volunteers using an anthropomorphic phantom for calibration. (June 2013)
- Record Type:
- Journal Article
- Title:
- Cross-camera comparison of ROI-based semi-quantitative 123I-IBZM SPECT data in healthy volunteers using an anthropomorphic phantom for calibration. (June 2013)
- Main Title:
- Cross-camera comparison of ROI-based semi-quantitative 123I-IBZM SPECT data in healthy volunteers using an anthropomorphic phantom for calibration
- Authors:
- Mo, Susanna Jakobson
Larsson, Anne
Johansson, Lennart
Stenlund, Hans
Forsgren, Lars
Riklund, Katrine - Abstract:
- Background: In 123 I-Iolopride (IBZM) SPECT reference values may diverge between camera systems. If multicenter pooling of normal material databases is needed, differences in measured semi-quantitative data due to equipment performance and reconstruction parameters have to be investigated in each instance to determine the comparability. Purpose: To explore the differences in 123 I-IBZM measured uptake ratios between two different gamma cameras in healthy controls, the intra-rater reproducibility of the image evaluation method and the possibility to equalize uptake ratios by calibration through an anthropomorphic phantom. Material and Methods: Differences in ROI-based semi-quantitative data from two different gamma camera systems, the three-headed brain dedicated Neurocam and the two-headed multipurpose hybrid system Infinia Hawkeye, were studied using image data from a group of healthy volunteers and an anthropomorphic brain-phantom scanned with both cameras. Several reconstruction methods and corrections were applied. To test the reliability of the ROI method, the intra-observer reproducibility was determined for the ROI method in this study. Results: The ROI method had a high reliability. Differences in mean measured uptake 123 I-IBZM ratios in healthy controls varied between 2.9% and 6.5% depending on reconstruction and correction for attenuation and scatter. After calibration, the differences decreased. There were no statistically significant differences betweenBackground: In 123 I-Iolopride (IBZM) SPECT reference values may diverge between camera systems. If multicenter pooling of normal material databases is needed, differences in measured semi-quantitative data due to equipment performance and reconstruction parameters have to be investigated in each instance to determine the comparability. Purpose: To explore the differences in 123 I-IBZM measured uptake ratios between two different gamma cameras in healthy controls, the intra-rater reproducibility of the image evaluation method and the possibility to equalize uptake ratios by calibration through an anthropomorphic phantom. Material and Methods: Differences in ROI-based semi-quantitative data from two different gamma camera systems, the three-headed brain dedicated Neurocam and the two-headed multipurpose hybrid system Infinia Hawkeye, were studied using image data from a group of healthy volunteers and an anthropomorphic brain-phantom scanned with both cameras. Several reconstruction methods and corrections were applied. To test the reliability of the ROI method, the intra-observer reproducibility was determined for the ROI method in this study. Results: The ROI method had a high reliability. Differences in mean measured uptake 123 I-IBZM ratios in healthy controls varied between 2.9% and 6.5% depending on reconstruction and correction for attenuation and scatter. After calibration, the differences decreased. There were no statistically significant differences between corrected ratios from the two camera systems in the study when images were reconstructed with attenuation correction. Conclusion: The conformity of uptake ratios in attenuation corrected 123 I-IBZM images derived from the two different cameras was improved by using an anthropomorphic phantom for calibration. … (more)
- Is Part Of:
- Acta radiologica. Volume 54:Number 5(2013)
- Journal:
- Acta radiologica
- Issue:
- Volume 54:Number 5(2013)
- Issue Display:
- Volume 54, Issue 5 (2013)
- Year:
- 2013
- Volume:
- 54
- Issue:
- 5
- Issue Sort Value:
- 2013-0054-0005-0000
- Page Start:
- 549
- Page End:
- 556
- Publication Date:
- 2013-06
- Subjects:
- SPECT -- IBZM -- radionuclide imaging -- imaging phantom -- Parkinsonian disorders -- healthy volunteers
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1177/0284185113477392 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24406.xml