A proposed protocol for acceptance and constancy control of computed tomography systems: A Nordic Association for Clinical Physics (NACP) work group report. (March 2013)
- Record Type:
- Journal Article
- Title:
- A proposed protocol for acceptance and constancy control of computed tomography systems: A Nordic Association for Clinical Physics (NACP) work group report. (March 2013)
- Main Title:
- A proposed protocol for acceptance and constancy control of computed tomography systems: A Nordic Association for Clinical Physics (NACP) work group report
- Authors:
- Kuttner, Samuel
Bujila, Robert
Kortesniemi, Mika
Andersson, Henrik
Kull, Love
Østerås, Bjørn Helge
Thygesen, Jesper
Tarp, Ivanka Sojat - Abstract:
- Background: Quality assurance (QA) of computed tomography (CT) systems is one of the routine tasks for medical physicists in the Nordic countries. However, standardized QA protocols do not yet exist and the QA methods, as well as the applied tolerance levels, vary in scope and extent at different hospitals. Purpose: To propose a standardized protocol for acceptance and constancy testing of CT scanners in the Nordic Region. Material and Methods: Following a Nordic Association for Clinical Physics (NACP) initiative, a group of medical physicists, with representatives from four Nordic countries, was formed. Based on international literature and practical experience within the group, a comprehensive standardized test protocol was developed. Results: The proposed protocol includes tests related to the mechanical functionality, X-ray tube, detector, and image quality for CT scanners. For each test, recommendations regarding the purpose, equipment needed, an outline of the test method, the measured parameter, tolerance levels, and the testing frequency are stated. In addition, a number of optional tests are briefly discussed that may provide further information about the CT system. Conclusion: Based on international references and medical physicists' practical experiences, a comprehensive QA protocol for CT systems is proposed, including both acceptance and constancy tests. The protocol may serve as a reference for medical physicists in the Nordic countries.
- Is Part Of:
- Acta radiologica. Volume 54:Number 2(2013)
- Journal:
- Acta radiologica
- Issue:
- Volume 54:Number 2(2013)
- Issue Display:
- Volume 54, Issue 2 (2013)
- Year:
- 2013
- Volume:
- 54
- Issue:
- 2
- Issue Sort Value:
- 2013-0054-0002-0000
- Page Start:
- 188
- Page End:
- 198
- Publication Date:
- 2013-03
- Subjects:
- CT -- QA/QC -- physics -- acceptance testing -- technical aspects -- equipment
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1258/ar.2012.120254 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24390.xml