FC 052ACID-BASE BALANCE DURING IN-SERIES EXTRACORPOREAL CARBON DIOXIDE REMOVAL AND CONTINUOUS VENOVENOUS HEMOFILTRATION: PREDICTIONS FROM A MATHEMATICAL MODEL. (29th May 2021)
- Record Type:
- Journal Article
- Title:
- FC 052ACID-BASE BALANCE DURING IN-SERIES EXTRACORPOREAL CARBON DIOXIDE REMOVAL AND CONTINUOUS VENOVENOUS HEMOFILTRATION: PREDICTIONS FROM A MATHEMATICAL MODEL. (29th May 2021)
- Main Title:
- FC 052ACID-BASE BALANCE DURING IN-SERIES EXTRACORPOREAL CARBON DIOXIDE REMOVAL AND CONTINUOUS VENOVENOUS HEMOFILTRATION: PREDICTIONS FROM A MATHEMATICAL MODEL
- Authors:
- Leypoldt, John (Ken)
Kurz, Joerg
Echeverri, Jorge
Storr, Markus
Harenski, Kai - Abstract:
- Abstract: Background and Aims: Critically ill acute kidney injury (AKI) patients may require treatment by extracorporeal carbon dioxide removal (ECCO2 R) devices to allow protective or ultraprotective mechanical ventilation and avoid hypercapnic acidosis. Continuous venovenous hemofiltration (CVVH) and ECCO2 R devices can be arranged in series to form a single extracorporeal circuit; such a circuit has been proposed to be optimal, based carbon dioxide removal efficacy, if the ECCO2 R device is placed proximal to the CVVH device (Allardet-Servent et al, Crit Care Med 43:2570-2581, 2015). Method: We developed a mathematical model of whole-body, acid-base balance during extracorporeal therapy using in-series ECCO2 R and CVVH devices for treatment of mechanically ventilated AKI patients. Equilibrium acid-base chemistry in blood was assumed as reported previously (Rees and Andreassen, Crit Rev Biomed Eng 33:209-264, 2005). Published clinical data from Allardet-Servent et al of mechanically ventilated (6 mL/kg predicted body weight or PBW) AKI patients treated by CVVH without ECCO2 R were used to adjust model parameters to fit plasma levels of arterial partial pressure of carbon dioxide (PaCO2 ) and arterial plasma bicarbonate concentration ([HCO3 ]). The effects of applying ECCO2 R at an unchanged tidal volume and a reduced tidal volume (4 mL/kg PBW) on PaCO2 and [HCO3 ] were then simulated assuming carbon dioxide removal rates from the ECCO2 R device measured in the clinicalAbstract: Background and Aims: Critically ill acute kidney injury (AKI) patients may require treatment by extracorporeal carbon dioxide removal (ECCO2 R) devices to allow protective or ultraprotective mechanical ventilation and avoid hypercapnic acidosis. Continuous venovenous hemofiltration (CVVH) and ECCO2 R devices can be arranged in series to form a single extracorporeal circuit; such a circuit has been proposed to be optimal, based carbon dioxide removal efficacy, if the ECCO2 R device is placed proximal to the CVVH device (Allardet-Servent et al, Crit Care Med 43:2570-2581, 2015). Method: We developed a mathematical model of whole-body, acid-base balance during extracorporeal therapy using in-series ECCO2 R and CVVH devices for treatment of mechanically ventilated AKI patients. Equilibrium acid-base chemistry in blood was assumed as reported previously (Rees and Andreassen, Crit Rev Biomed Eng 33:209-264, 2005). Published clinical data from Allardet-Servent et al of mechanically ventilated (6 mL/kg predicted body weight or PBW) AKI patients treated by CVVH without ECCO2 R were used to adjust model parameters to fit plasma levels of arterial partial pressure of carbon dioxide (PaCO2 ) and arterial plasma bicarbonate concentration ([HCO3 ]). The effects of applying ECCO2 R at an unchanged tidal volume and a reduced tidal volume (4 mL/kg PBW) on PaCO2 and [HCO3 ] were then simulated assuming carbon dioxide removal rates from the ECCO2 R device measured in the clinical study (91 mL of CO2 /min when ECCO2 R was proximal and 72 mL of CO2 /min when CVVH was proximal). Results: Agreement of model predictions with the clinical data was good, and model predictions were relatively independent of the in-series position of the devices (see Table ). Total carbon dioxide removal from the CVVH device via ultrafiltration predicted by the model was lower after applying ECCO2 R at both the unchanged tidal volume (25 mL of CO2 /min when ECCO2 R was proximal and 39 mL of CO2 /min when CVVH was proximal) and the reduced tidal volume (30 mL of CO2 /min when ECCO2 R was proximal and 44 mL of CO2 /min when CVVH was proximal). The reduced removal of total carbon dioxide via ultrafiltration when ECCO2 R was proximal resulted from the lower total carbon dioxide concentration in blood entering the CVVH device. Thus, independent of the in-series position of the devices, the magnitude of this difference in total carbon dioxide removal by the CVVH device (14 mL of CO2 /min) approximately cancels out the relative greater efficacy of the ECCO2R device (19 mL of CO2 /min). Conclusion: The described mathematical model has quantitative accuracy. It suggests that overall acid-base balance when using ECCO2 R and CVVH devices in a single, combined extracorporeal circuit will be similar, independent of their in-series position. … (more)
- Is Part Of:
- Nephrology dialysis transplantation. Volume 36(2021)Supplement 1
- Journal:
- Nephrology dialysis transplantation
- Issue:
- Volume 36(2021)Supplement 1
- Issue Display:
- Volume 36, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 36
- Issue:
- 1
- Issue Sort Value:
- 2021-0036-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-05-29
- Subjects:
- Nephrology -- Periodicals
Hemodialysis -- Periodicals
Kidneys -- Transplantation -- Periodicals
Hemodialysis
Kidneys -- Transplantation
Nephrology
Periodicals
616.61 - Journal URLs:
- http://ndt.oxfordjournals.org/ ↗
http://www.oup.co.uk/ndt/ ↗
http://ukcatalogue.oup.com/ ↗
http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=0931-0509;screen=info;ECOIP ↗ - DOI:
- 10.1093/ndt/gfab140.004 ↗
- Languages:
- English
- ISSNs:
- 0931-0509
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6075.685300
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24345.xml