Comparison of carbapenem minimum inhibitory concentrations of Oxacillin-48-like Klebsiella pneumoniae by Sensititre, Vitek 2, MicroScan, and Etest. (December 2022)
- Record Type:
- Journal Article
- Title:
- Comparison of carbapenem minimum inhibitory concentrations of Oxacillin-48-like Klebsiella pneumoniae by Sensititre, Vitek 2, MicroScan, and Etest. (December 2022)
- Main Title:
- Comparison of carbapenem minimum inhibitory concentrations of Oxacillin-48-like Klebsiella pneumoniae by Sensititre, Vitek 2, MicroScan, and Etest
- Authors:
- Nana, Trusha
Perovic, Olga
Chibabhai, Vindana - Abstract:
- Abstract: Objectives: The aim of this laboratory-based study was to compare carbapenem MICs yielded by Sensititre, Vitek 2, MicroScan WalkAway plus and Etest for Oxacillin (OXA)-48-like Klebsiella pneumoniae isolates. Methods: Analysis was performed for categorical agreement for ertapenem, meropenem, and imipenem, and the proportion of isolates with MICs ≤8μg/mL and the MIC50/MIC90 for meropenem and imipenem, from a convenience sample of 82 deduplicated blood culture OXA-48-like K. pneumoniae isolates. Results: The proportion of isolates testing susceptible to ertapenem by Etest (19/82, 23.1%) differed from Sensititre/Vitek (0/82) and MicroScan (2/82, 2.4%) (p < 0.001 for all). For meropenem, the proportion of isolates susceptible by Etest (31/82, 37.8%) differed from Sensititre/Vitek (16/82, 19.5%) (p = 0.015). There was variation in the proportion of isolates that tested imipenem susceptible when comparing Sensititre (9/82, 11%) and Vitek (8/82, 9.8%) to MicroScan (27/82, 32.9%), p = 0.001 and p < 0.001, respectively, Sensititre and Vitek to Etest (45/82, 54.9%), p < 0.001 for both, and MicroScan to Etest, p = 0.007. The proportion of isolates with meropenem MICs ≤8μg/mL with Sensititre and Vitek differed significantly from Etest, 58.5% and 85.4%, respectively, p < 0.001. A 2-fold difference between the Sensititre and Vitek meropenem and imipenem MIC at which ≥50% of isolates were inhibited compared to the MicroScan, and a 4-fold difference compared to Etest, was present.Abstract: Objectives: The aim of this laboratory-based study was to compare carbapenem MICs yielded by Sensititre, Vitek 2, MicroScan WalkAway plus and Etest for Oxacillin (OXA)-48-like Klebsiella pneumoniae isolates. Methods: Analysis was performed for categorical agreement for ertapenem, meropenem, and imipenem, and the proportion of isolates with MICs ≤8μg/mL and the MIC50/MIC90 for meropenem and imipenem, from a convenience sample of 82 deduplicated blood culture OXA-48-like K. pneumoniae isolates. Results: The proportion of isolates testing susceptible to ertapenem by Etest (19/82, 23.1%) differed from Sensititre/Vitek (0/82) and MicroScan (2/82, 2.4%) (p < 0.001 for all). For meropenem, the proportion of isolates susceptible by Etest (31/82, 37.8%) differed from Sensititre/Vitek (16/82, 19.5%) (p = 0.015). There was variation in the proportion of isolates that tested imipenem susceptible when comparing Sensititre (9/82, 11%) and Vitek (8/82, 9.8%) to MicroScan (27/82, 32.9%), p = 0.001 and p < 0.001, respectively, Sensititre and Vitek to Etest (45/82, 54.9%), p < 0.001 for both, and MicroScan to Etest, p = 0.007. The proportion of isolates with meropenem MICs ≤8μg/mL with Sensititre and Vitek differed significantly from Etest, 58.5% and 85.4%, respectively, p < 0.001. A 2-fold difference between the Sensititre and Vitek meropenem and imipenem MIC at which ≥50% of isolates were inhibited compared to the MicroScan, and a 4-fold difference compared to Etest, was present. Conclusions: Substantial variability in carbapenem MICs for OXA-48-like carbapenemase-producing Enterobacterales isolates by the four methods was demonstrated. Performance characteristics verification of MIC methods in use for the predominant carbapenemase-producing Enterobacterales type is required by laboratories to optimize the accuracy of carbapenem reporting. Graphical abstract: Image 1 … (more)
- Is Part Of:
- Clinical microbiology and infection. Volume 28:Number 12(2022)
- Journal:
- Clinical microbiology and infection
- Issue:
- Volume 28:Number 12(2022)
- Issue Display:
- Volume 28, Issue 12 (2022)
- Year:
- 2022
- Volume:
- 28
- Issue:
- 12
- Issue Sort Value:
- 2022-0028-0012-0000
- Page Start:
- 1650.e1
- Page End:
- 1650.e5
- Publication Date:
- 2022-12
- Subjects:
- Accuracy -- Carbapenemase-producing Enterobacterales -- Commercial antimicrobial susceptibility testing methods -- Etest -- Klebsiella pneumoniae -- MicroScan -- Minimum inhibitory concentration -- OXA-48-like -- Sensititre -- Vitek 2
Medical microbiology -- Periodicals
Diagnostic microbiology -- Periodicals
Communicable diseases -- Periodicals
Infection -- Periodicals
616.01 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1469-0691 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1016/j.cmi.2022.06.023 ↗
- Languages:
- English
- ISSNs:
- 1198-743X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.305520
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24319.xml