A new stationary grid, with grid lines aligned to pixel lines with submicron‐order precision, to suppress grid artifacts. Issue 9 (28th July 2021)
- Record Type:
- Journal Article
- Title:
- A new stationary grid, with grid lines aligned to pixel lines with submicron‐order precision, to suppress grid artifacts. Issue 9 (28th July 2021)
- Main Title:
- A new stationary grid, with grid lines aligned to pixel lines with submicron‐order precision, to suppress grid artifacts
- Authors:
- Kawashima, Hiroki
Ichikawa, Katsuhiro
Iida, Yasuko - Abstract:
- Abstract: Purpose: We have developed a new stationary grid named a pixel‐aligned grid (PA grid), in which the grid lines are aligned to the pixel lines with submicron‐order precision. Further, we have evaluated its performance relative to that of a conventional grid combined with grid‐line removal (GLR) processing. Methods: A flat‐panel detector system of an indirect type, with a pixel pitch of 150 μm, was employed. Four PA grids having a grid ratio of 6:1 associated with abdominal bedside radiography, with the grid‐line pitch (GP) varied around the target value of 150 μm, were produced. Blank images were obtained with four PA grids for measuring the period and amplitude of the grid artifact. In performance evaluation, acrylic and anthropomorphic abdominal phantom images were used with the PA grid, a conventional grid (40 lines/cm, grid ratio 6:1), and no grids. The grid artifacts were evaluated by power spectrum (PS) analysis. Also, the signal‐to‐noise ratio (SNR) improvement factor ( K SNR ) was measured. Results: Grid artifacts were hardly recognizable with PA grids with GP errors of 0.3 μm and 0.6 μm because of the prolonged grid artifact periods. The measured artifact amplitudes of these PA grids were less than 0.6%. Furthermore, the PA grids did not produce notable frequency peaks in PS. In contrast, the conventional grid without GLR processing produced two conspicuous peaks. With GLR processing, notable reductions in PS were observed around the two peak frequencies,Abstract: Purpose: We have developed a new stationary grid named a pixel‐aligned grid (PA grid), in which the grid lines are aligned to the pixel lines with submicron‐order precision. Further, we have evaluated its performance relative to that of a conventional grid combined with grid‐line removal (GLR) processing. Methods: A flat‐panel detector system of an indirect type, with a pixel pitch of 150 μm, was employed. Four PA grids having a grid ratio of 6:1 associated with abdominal bedside radiography, with the grid‐line pitch (GP) varied around the target value of 150 μm, were produced. Blank images were obtained with four PA grids for measuring the period and amplitude of the grid artifact. In performance evaluation, acrylic and anthropomorphic abdominal phantom images were used with the PA grid, a conventional grid (40 lines/cm, grid ratio 6:1), and no grids. The grid artifacts were evaluated by power spectrum (PS) analysis. Also, the signal‐to‐noise ratio (SNR) improvement factor ( K SNR ) was measured. Results: Grid artifacts were hardly recognizable with PA grids with GP errors of 0.3 μm and 0.6 μm because of the prolonged grid artifact periods. The measured artifact amplitudes of these PA grids were less than 0.6%. Furthermore, the PA grids did not produce notable frequency peaks in PS. In contrast, the conventional grid without GLR processing produced two conspicuous peaks. With GLR processing, notable reductions in PS were observed around the two peak frequencies, which caused blurring in bone structures. For the acrylic thickness of 20 cm, the K SNR s for the PA grid were around 1.4, suggesting some SNR improvement in abdominal bedside radiography. Conclusion: The present study has demonstrated that PA grids with their grid‐line pitches close to the pixel‐line pitch within errors of 0.6 μm produce grid artifact‐free images without any signal losses. Thus, the proposed PA grid will prove to be effective and useful in various clinical applications. … (more)
- Is Part Of:
- Medical physics. Volume 48:Issue 9(2021)
- Journal:
- Medical physics
- Issue:
- Volume 48:Issue 9(2021)
- Issue Display:
- Volume 48, Issue 9 (2021)
- Year:
- 2021
- Volume:
- 48
- Issue:
- 9
- Issue Sort Value:
- 2021-0048-0009-0000
- Page Start:
- 4935
- Page End:
- 4943
- Publication Date:
- 2021-07-28
- Subjects:
- anti‐scatter grid -- digital radiography -- grid artifacts -- image quality -- scatter fraction
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.15099 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24280.xml